222
L. Karabulut et al. / Chemical Physics Letters 503 (2011) 220–225
objective. The photographs of the equilibrated grain boundary
groove shapes and the graticule were superimposed on one an-
other using Adobe PhotoShop 8.0 version software so that accurate
measurement of the groove coordinate points on the groove shapes
could be made.
full. The specimen was then placed in the radial heat flow
apparatus.
The specimen was heated from the centre using a single heating
wire (140–190 mm in length and 2.5 mm in diameter, Kanthal A–1)
in steps of 50 K up to 10 K below the melting temperature of the
alloy and the outside of the specimen was cooled to maintain a ra-
dial temperature gradient. To obtain a reliable value of thermal
conductivity in the thermal conductivity measurement, a larger ra-
dial temperature gradient is desired. For this purpose, the gap be-
tween the cooling jacket and the specimen was filled with free
running sand or graphite dust and the outside of the specimen
was kept at 283 K using a heating/refrigerating circulating bath.
The length of central heating wire was chosen to be 10 mm longer
than the length of specimen to make isotherms parallel to the ver-
tical axis.
2.3. Geometrical correction for the groove coordinates
The coordinates of the cusp, x, y should be measured using the
coordinates x, y, z where the x axis is parallel to the solid–liquid
interface, the y axis is normal to the solid–liquid interface and
the z axis lies at the base of the grain boundary groove. Marasßlı
and Hunt [14] devised a geometrical method to make appropriate
corrections to the groove shapes and the details of the geometrical
method are given in Ref. [14].
The coordinates of equilibrated grain boundary groove shapes
were measured with an optical microscope to an accuracy of
10 lm by following Maraßslı and Hunt’s geometrical method so
that appropriate corrections to the shape of the grooves could be
deduced [14]. The uncertainty in the measurements of equilibrated
grain boundary groove coordinates is 0.1%.
The specimen was kept at steady-state condition for at least
two hours for a setting temperature. At steady state, the total input
power and the stationary thermocouple temperatures were re-
corded with a Hewlett Packard 34401 type multimeter and a Pico
TC–08 data–logger. The temperatures on the different parts of the
specimen were measured with mineral insulated metal sheathed,
0.5 mm in diameter K type thermocouples. The zero or minimum
vertical temperature gradient is desired in thermal conductivity
measurements. The vertical temperature for each setting was tried
to be made as parallel as possible to the vertical axis by moving the
central heater up and down. After all desired power settings and
temperature measurements had been completed during the heat-
ing procedure, the cooling procedure was started in same steps
down to room temperature.
2.4. Measurements of the thermal conductivity of solid and liquid
phases
The thermal conductivity ratio of equilibrated eutectic liquid
phase (Pb–17.5 at.% Sb) to solid Pb solution (Pb–5.8 at.% Sb) phase,
R = KL(eutectic liquid)/KS(Pb) must be known or measured to evaluate
the Gibbs–Thomson coefficient with the present numerical meth-
od. The radial heat flow method is an ideal technique for measur-
ing the thermal conductivity of the solid phase. The thermal
conductivity of the solid Pb solution phase is also needed to eval-
uate the temperature gradient in the solid phases. In the radial heat
flow method, a cylindrical sample was heated by using a single
heating wire along the axis at the centre of the sample and the
sample was kept in a very stable temperature gradient for a period
to achieve a steady-state condition. At the steady–state condition,
the temperature gradient in the cylindrical specimen is given by
Fourier’s law,
Then the sample was removed from the furnace and cut trans-
versely near to the measurement points, after that the specimen
was ground and polished for the measurements of r1 and r2. The
positions of the thermocouples were then photographed with an
Olympus DP12 CCD digital camera placed in conjunction with an
Olympus BX51 type light optical microscope.
A
graticule
(200 Â 0.01 = 2 mm) was also photographed using the same objec-
tive. The photographs of the positions of the thermocouples and
the graticule were superimposed on one another using Adobe
PhotoShop 8.0 software so that accurate measurement of the dis-
tances of stationary thermocouples could be made to an accuracy
dT
dr
Q
of 10 lm. The transverse and longitudinal sections of the speci-
GS ¼
¼ À
ð4Þ
2pr‘KS
men were examined for the porosity, crack and casting defects to
make sure that these would not introduce any error to the mea-
surements. The experimental value of a0 for solid Pb solution and
Pb–Sb eutectic solid were 1.470, 1.487 mÀ1, respectively in the
present work.
where Q is the total input power, r is the distance of the solid–liquid
interface to the centre of the sample, ‘ is the length of the heating
wire which is constant and KS is the thermal conductivity of the so-
lid phase. Integration of the Eq. (4) gives
The thermal conductivities of solid Pb solution and eutectic so-
lid versus temperature are shown in Figure 2. A comparison of
thermal conductivities of solid Pb solution and Pb–Sb eutectic solid
with the thermal conductivity of Pb [21] and Sb [22] are also given
in Figure 2. The values of KS for the solid Pb solution and Pb–Sb eu-
tectic solid at the eutectic melting temperature were obtained to
be 28.7 and 26.8 W/Km, respectively by extrapolating to the eutec-
tic temperature as shown in Figure 2. The thermal conductivities of
solid and liquid phases for solid Pb solution and Pb–Sb eutectic so-
lid and their ratios are given in Table 1.
Q
KS ¼ a0
ð5Þ
T1 À T2
where a0 ¼ lnðr2=r1Þ=2
p
‘ is an experimental constant, r1 and r2
(r2 > r1) are fixed distances from the central axis of the specimen,
T1 and T2 are the temperatures at the fixed positions, r1 and r2.
Eq. (5) could be used to obtain the thermal conductivity of the solid
phase by measuring the difference in temperature between two
fixed points for a given power level provided that the vertical tem-
perature variation are minimum or zero.
The thermal conductivities of solid Pb solution (Pb–5.8 at.% Sb)
and eutectic solid (Pb–17.5 at.% Sb) were measured with the radial
heat flow apparatus. Sufficient amount of metallic materials were
melted to produce an ingot of approximately 100 mm in length
and 30 mm in diameter in a vacuum furnace by using 99.99% pure
Pb and 99.99% pure Sb. After stirring, the molten metallic alloy was
poured into a graphite crucible held in a specially constructed hot
filling furnace at approximately 100 K above the melting tempera-
ture of alloy. The molten metallic alloy was then directionally fro-
zen from bottom to top to ensure that the crucible was completely
It is not possible to measure the thermal conductivity of the li-
quid phase with the radial heat flow apparatus since a thick liquid
layer (10 mm) is required. A layer of this size would certainly have
led to convection. If the thermal conductivity ratio of the liquid
phase to the solid phase is known and the thermal conductivity
of the solid phase is measured at the melting temperature, the
thermal conductivity of the liquid phase can then be evaluated.
The thermal conductivity ratio can be obtained during directional
growth with a Bridgman type growth apparatus. The detail of the
experimental procedure was given in Refs. [13–17].