Angewandte
Chemie
Table 1: Electrochemical detection of 19 influenza strains and validation
with rRT–PCR and cell-culture plaque assays.
resulted in the release of glucose (Figure 2a, samples C, D,
and E). The cleavage was confirmed by mass spectral analysis.
The mass spectra of the control where there is no virus or NA
revealed a peak at m/z 494.1497 (M+Na, positive ion)
corresponding to uncleaved SG1. A new peak emerges at
m/z 181.0711 (M+1, positive ion) corresponding to glucose
when virus or NA is added.
To test for drug susceptibility, we premixed FDA-
approved NA inhibitors Zanamivir or Oseltamivir with the
virus or NA for 30 min before introducing SG1. If the strains
are not resistant to the antivirals, they are expected to block
the action of NA and a signal for glucose should not be
detected. As seen in Figure 2b, the three strains and NA are
completely inhibited by the antivirals. Thus, we determined
drug susceptibility of these strains within a sample results in
a test time of less than 2 hours. This is highly significant
because drug susceptibility for influenza viruses using geno-
typic and/or phenotypic methods requires sophisticated
instruments, trained personnel, and several hours to
complete.[17]
To differentiate between bacteria/human NA that are also
expected to cleave SG1, we exploited known differences in
the binding pocket of NAs. Bacteria/human NA have
a smaller binding pocket and cannot accommodate larger
groups at the 4-position of sialic acid, as confirmed by X-ray
structures and functional assays with Zanamivir and Oselta-
mivir.[10] We, and others, have exploited this feature to
develop inhibitors[10,18] and substrates[19] that are highly
specific for influenza NA. As expected, bacterial NA
(BNA) from Clostridium perfringens and viral NA cleave
SG1 (Figure 2c, sample BNA). However, when Zanamivir
was premixed with both NAs, only BNA cleaves SG1 and not
the viral NA (VNA) because the antiviral is specific for VNA
(Figure 2c). Therefore, we can distinguish between BNA and
viral NA using the antivirals. An alternate approach is to
introduce larger groups at the 4-position of sialic acid to make
the substrate highly specific for viral NA instead of using
Zanamivir; the syntheses of these next-generation substrates
are ongoing.
Next, we were interested in determining if nasal or throat
swabs, the standard source of clinical samples for influenza
viruses, have a background level of glucose. We found glucose
is absent in the nose or throat by testing samples obtained
from healthy human volunteers. Glucose is released when the
samples are spiked with known concentrations of the virus
(Figure 2d). Thus, this assay could be used to test influenza in
nasal and/or throat swabs.
Influenza strains
Plaque
Ct[b] i [10À8 A][c]
Assay[a]
No virus + SG1
No glucose
b-Galactosidase
a-Mannosidase
n/a
n/a
n/a
n/a
n/a
n/a
n/a
n/a
3.6Æ2.2
3.4Æ1.4
4.2Æ1.9
4.7Æ1.6
Glucose (1 mm)
n/a
n/a 124.9Æ2.3
A/Wilson-Smith/1933 (H1N1)
A/PuertoRico/8/1934 (H1N1)
A/HongKong/8/1968 (H3N2)
A/Aichi/2/1968 (H3N2)
A/Beijing/262/1995 (H1N1)
A/Nanchang/933/1995 (H3N2)
A/Sydney/5/1997 (H3N2)
A/NewCaledonia/20/1999 (H1N1)
A/SolomonIslands/3/2006 (H1N1)
A/Uruguay/716/2007 (H3N2)
A/Brisbane/59/2007 (H1N1)
A/Brisbane/10/2007 (H3N2)
A/California/07/2009 (H1N1)
A/New York/18/2009 (H1N1)
A/San Diego/1/2009 (H1N1)
A/Wisconsin/629-D02452/2009
(H1N1)
4.2106
1.4104
1.5105
1.0105
3.5105
7.0105
8.0103
4.4107
1.1109
1.2107
2.5107
2.2106
3.6106
3.5105
1.2105
6.0104
18
15
22
15
99.5Æ4.4
13.4Æ2.0
64.4Æ5.2
31.3Æ5.3
21 118.1Æ11.6
18
15.1Æ2.8
29 115.7Æ4.3
12
11
17.7Æ2.6
79.7Æ5.0
20 132.4Æ6.3
12 106.2Æ5.1
21
13
17
78.6Æ4.9
13.4Æ1.5
74.4Æ5.6
19 108.5Æ12.0
13
98.9Æ5.7
A/Wisconsin/15/2009 (H3N2)
A/Brownsville/31H/2009 (H1N1)
A/Victoria/361/2011 (H3N2)
5.0103
4.0103
7.0106
26 107.8Æ5.9
22
26
91.7Æ9.7
94.5Æ2.3
[a] Reported in pfu mLÀ1. [b] rRT–PCR was performed using 100 mL of
virus to measure Ct, that is, the cycle number at which fluorescence is
above the threshold (background). Ct is inversely proportional to the
number of amplicons. [c] Measured by electrochemical assay. 100 mL of
virus was mixed with 100 mL of PBS buffer with SG1 for 1 hour at 378C.
Glucose concentration was determined using 20 mL of this solution.
2’-(4-methylumbelliferyl)-a-d-N-acetylneuraminic
acid
(MUNANA; Figure S4). As expected, the rate of cleavage
of A/PuertoRico/8/1934 (H1N1), A/California/07/2009
(H1N1), and A/New Caledonia/20/1999 (H1N1) strains are
slower than the A/Beijing/262/1995 (H1N1) strain. We note
that, despite variations in printed electrodes from different
manufacturer or different batches from the same manufac-
turer, the assay detects all strains. We also determined the
analytical sensitivity using one of the strains using this
rudimentary setup (Figure S3). The limit of detection and
range is 102 and 102–108 pfu, respectively. As multiple studies
have reported that patients (n > 50) suffering from influenza
typically harbor 103–108 pfumLÀ1 in the nose/throat,[20] this
assay could be useful for rapid detection in a primary-care
setting.
To summarize, we have developed an electrochemical
assay that releases glucose upon introduction of influenza
viruses. We successfully detected 19 influenza strains. The
assay can be used to measure drug susceptibility rapidly,
a significant advantage over current genotypic and pheno-
typic methods that take time, resources, and a laboratory
environment.[17] The assay can be integrated into current
glucose meters by repurposing the instruments to test nasal or
throat swabs for influenza. As glucose meters with disposable
test strips are user friendly, ubiquitous, and inexpensive, this
To improve assay performance, we used disposable
printed electrodes (CH instruments, Austin, Texas) for the
next set of experiments (Figure S3). This experimental setup
requires only 20 mL of solution, similar to commercial
disposable test strips used in blood glucose meters. We
obtained 19 H1N1 and H3N2 influenza strains that spanned
eighty years, from 1933 to 2011, including strains from the
most recent 2009 pandemic. All strains were detected in one
hour, which demonstrates broad specificity. These results
were validated using rRT–PCR and plaque assays (Table 1).
Some strains cleaved SG1 slowly; we corroborated the results
by measuring the slow release using a fluorescent substrate,
Angew. Chem. Int. Ed. 2015, 54, 5929 –5932
ꢀ 2015 Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim
5931