896
X.-H. Sun et al. / J. Chem. Thermodynamics 36 (2004) 895–899
2. Experimental
in the temperature range of (78 to 364) K to eliminate
the heat loss owing to the gas convection, and a small
amount of helium gas was introduced through a length
of copper capillary at the centre of the upper cover into
the cell to improve the heat transfer of the whole sample
cell. The sample cell was sealed with the tin solder after
the copper capillary was pinched off from the tube end.
Two adiabatic shields surrounded the sample cell in turn
and a vacuum can was immersed in liquid nitrogen. The
two adiabatic shields were made of chromium-plated
copper and equipped with manganin heating wires.
Two sets of six-junction chromel-contantan (Ni 55%,
Cu 45%) thermocouples were used to measure the tem-
perature differences between the sample cell and the in-
ner adiabatic shield and between the inner and the
outer adiabatic shields. The temperatures of the two
shields were controlled separately and automatically
with two units of auto-adiabatic controller. When the
2.1. Sample preparation and characterization
2.1.1. Equipment and reagents
Pyrimethanil laurate was made according to the
method reported in literature [5], pyrimethanil (purified
3 times, the purity was 0.9980 mole fraction, m.p. 96.3
ꢁC [6]), ethanol, lauric acid, were of analytical grade.
PE-2400 elemental analyzer, BRUKER EQUNINOX-
1
55 IR spectral meter, Varian Inova-400 H NMR spec-
trum, Agilent gas chromatograph/Hewlett Packard mass
spectrometry and SMP3 melting point apparatus were
applied to characterize the structure of the compound.
2.1.2. The scheme of the formation reaction of pyrimeth-
anil laurate
CH
CH3
N
3
N
N
EtOH
.
+
NH
CH (CH ) COOH
3 2 10
NH
CH (CH ) COOH
3
2 10
N
CH
CH3
3
2.1.3. Experimental procedures
temperature in the sample cell increases due to heating,
the thermocouples measure the temperature differences.
This signal is used to control the heaters distributed on
the walls of the inner and outer shields, respectively.
Both shields were heated under the control of the signal
and kept at the same temperature as that of the sample
cell. In this way, the heat loss caused by the radiation is
greatly reduced. The miniature platinum resistance ther-
mometer (IPRT No. 2, produced by Shanghai Institute
of Industrial Automatic Meters, 16 mm in length, 1.6
mm in diameter and a nominal resistance of 100 X)
was applied to measure the temperature of the sample.
The thermometer was calibrated on the basis of ITS-
90 by the Station of Low-temperature Metrology and
Measurements, Academia Sinica.
The sample was heated using the standard discrete
heating method and the temperature of the sample
was alternatively measured. The heating duration was
10 min, the equilibrium time of each temperature point
is 5 min, and the temperature drift rates of the sample
cell measured in an equilibrium period were usually
within (10ꢀ3 to 10ꢀ4) K Æ minꢀ1. During the heat-ca-
pacity measurements, the temperature difference be-
tween the inner adiabatic shield and the sample cell
was automatically kept within 10ꢀ3 K Æ minꢀ1 in order
to obtain a satisfactory adiabatic effect and corre-
sponding equilibrium temperature were corrected for
heat loss [7,8].
A solution of pyrimethanil and lauric acid in anhy-
drous alcohol was stirred for 1 h at room temperature.
The precipitate formed was filtered and recrystallized
from anhydrous alcohol to yield compound. The final
product pyrimethanil laurate was a colorless crystal.
Its mass fraction purity was 0.9957 (GC).
2.1.4. Analysis result of pyrimethanil laurate
The melting point of the final product was deter-
mined to be T = (321.55 to 322.35) K. The results of ele-
mental analysis were: N: 10.51 (10.52); C: 72.08 (72.16);
H: 9.26 (9.34). 1H MNR (CDCl3) absorption peaks were
detected at d = 0.88 (t, 3 H), 1.29 to 1.32 (m, 18 H), 1.63
to 1.71 (m, 3 H), 2.38 (s, 6 H), 6.46 (s, 1 H), 6.99 to 7.75
(m, 5 H), 8.92 (s, 1 H ) 1 · 10ꢀ6. IR (KBr disks) showed
characteristic absorption peaks at 3294 (NH), 2922
(CH), 2546 ðNHþ2 Þ, 1703 (C‚O), 1628 (Ph) cmꢀ1
.
2.1.5. Adiabatic calorimetry
The heat capacity measurements were made by an
adiabatic calorimetric system for small samples over
the temperature range between T = (78 and 340) K.
The construction of the calorimeter has been described
previously [5,6] in detail. It consists of a sample cell, a
platinum resistance thermometer, a heater, two (inner
and outer) adiabatic shields, two sets of differential ther-
mocouples, a vacuum can and a Dewar vessel. Liquid
nitrogen was used as the cooling medium. The evacu-
ated can or chamber was kept within 10ꢀ3 Pa vacuum
The mass of the sample loaded in the sample cell
amounted to 2.1856 g, which was equivalent to
0.00547 mol based on its molar mass of 399.57 g Æ molꢀ1
.