1
32
W. Wang et al. / Journal of Catalysis 352 (2017) 130–141
samples were denoted as 0%Fe-5%Cr, 5%Fe-5%Cr, 10%Fe-5%Cr or
rate of 80 mL/min), the samples were heated from room tempera-
ture to 378 K, keeping at 378 K for 10 min to remove the moisture
content, and then further heated up to 1173 K. Besides, all samples
were kept at 10 ±1 mg for avoiding heat transfer limitations.
2
0%Fe-5%Cr-PKU-1, respectively.
2.2. Catalyst characterizations
2
.2.7 Photoluminescence (PL) spectroscopy was used to detect
2
.2.1 Powder X-ray diffraction data were collected at room tem-
perature on a PANalytical X’pert powder diffractometer equipped
with a PIXcel 1D detector under Cu K radiation (k = 1.5406 Å).
OH radicals formed over Cr-PKU-1 catalyst using terephthalic acid
(TA) as a probe molecule. In a typical run, 40 mg of 10%Cr-PKU-1
was dispersed in a mixed aqueous solution containing 3 mmol
a
The operating voltage and current were 40 kV and 40 mA, respec-
tively. Le Bail fitting was performed using the TOPAS software
package [41]. In order to in-situ observe the phase transformation
under higher temperature, a high-temperature powder XRD exper-
iments were carried out on a Bruker D8 Advance diffractometer
3 2 2
TA, 20 mmol CHOL, 20 mL CH CN, and 40 mmol H O . After stirring
for a specified interval of time, the suspension was withdrawn and
then centrifuged to remove the powder. The upper and transparent
solution was used for the PL measurement with an excitation irra-
diation at 328 nm on a Hitachi F4600 fluorescence spectrometer.
PMT voltage was fixed to be 700 V, and the width of excitation
and emission slit were both set to 2.5 nm.
(
Cu Ka radiation, k = 1.5406 Å) with a computer-controlled fur-
nace. The sample was loaded on a platinum strip and heated from
room temperature to 1300 K in air at a heating rate of 5 K/min, and
the XRD data were collected after the temperature was stabilized
for 1 h at each chosen temperature.
2.2.8 UV–Vis absorption spectra were used to detect superoxide
Å
À
2
ion ( O ) formed over Cr-PKU-1 catalyst using nitroblue tetra-
zolium (NBT) as a probe molecule. In a typical run, 40 mg of 10%
Cr-PKU-1 was dispersed in a mixed aqueous solution containing
3 mmol NBT, 20 mmol CHOL, 20 mL CH CN, and 40 mmol H O .
3 2 2
After stirring for a specified interval of time, the suspension was
withdrawn and then centrifuged to remove the catalyst. The upper
solution was diluted for 20 times with DMSO and then used for
UV–Vis absorption measurement in a UV–Vis spectrometer (PUXI
TU-1810, China).
2
.2.2 X-ray photoelectron spectra (XPS) were acquired with UK
Kratos Axis Ultra spectrometer with Al K X-ray source operated at
5 kV and 15 mA. Kinetic energies of photoelectrons were mea-
a
1
sured using a hemispherical electron analyzer working at the con-
stant pass energy (40 eV). The pressure in the chamber was less
À9
than 5.0 Â 10 Torr. The XPS data were collected in increments
of 0.1 eV with dwell time of 500 ms. Electron binding energies
b
were calibrated against the C 1 s emission at E = 284.6 eV to cor-
rect the contact potential differences between the sample and
the spectrometer.
2.3. Catalytic dehydrogenation
2
.2.3 Cyclic voltammetry (CV) was performed to identify the
All organic solvents were purified by distillation method, other
substrates such as sec-alcohol were used as received. Catalytic
dehydrogenation of sec-alcohols was performed in a 50 mL
single-neck round-bottom flask immersed into a silicon oil bath
and stirred with a magnetic stirrer. In a standard run, 20 mmol
redox behavior of Cr-PKU-1. To prepare the testing electrode, a vis-
cous slurry containing 74 wt % Cr-PKU-1, 18.5 wt % acetylene
black, and 7.5 wt % PTFE was mixed and pressed onto a nickel foam
current collector. After that, the electrode was dried in vacuum and
maintained at 333 K for 24 h in prior to measurements. Electro-
chemical tests were examined on an electrochemical workstation
3
CHOL, 20 mL CH CN, 40 mg 10%Cr-PKU-1 catalyst and 40 mmol
2 2
H O
(30 wt %) were mixed in the flask. After heated at 353 K for
(
Autolab Pgstat302N, Switzerland). The working solution was pre-
pared by mixing CH CN (80 mL), dimethyl sulphoxide (DMSO)
0.5 mL) and 0.1 M KCl solution (10 mL). In a three-electrode cell,
8 h, reaction mixture was extracted with a syringe to transfer to
a centrifuge tube, and subsequently centrifuged to remove the
solid catalyst. Finally, the obtained transparent solution was ana-
lyzed qualitatively or quantitatively with chromatography-mass
spectrometry (GC–MS), gas chromatography (GC) and high perfor-
mance liquid chromatography (HPLC), respectively.
3
(
Pt foil, Ag/AgCl and the above loaded nickel foam were used as
the counter, reference and working electrodes, respectively. Cyclic
voltammetry curves were obtained in the potential range of
À1.0 ꢀ 0 V vs. Ag/AgCl by varying the scan rate from 5 to
GC–MS was performed on an Agilent 7890N gas chromatograph
coupled with a capillary column (DB-5: length, 30 m; inner diam-
eter, 0.25 mm) and mass spectrum system (Agilent 5975C). Helium
was used as both the carrier (1 mL/min) and make-up gas (40 mL/
min). GC–MS spectra were identified by comparing the m/z value of
each peak with the NIST Mass Spectral database.
GC was used to determine the concentration of the substrates
and products on a Techcomp GC 7800 gas chromatograph
equipped with a flame ionization detector (FID). A capillary column
(CP-43: length, 25 m; inner diameter, 0.53 mm; film thickness,
À1
1
00 mVÁs
.
2
.2.4 Sample morphology was examined with a field emission
scanning electron microscopy (JEOL, JSM-7800F) equipped with
an energy dispersive spectrometer (EDS) analyzer. Samples were
prepared by dispersing dry powder on double-sided conductive
adhesive tape followed by coating with a gold film. The recorded
SEM image shown in this work was at a magnification of 1300Â.
2
.2.5 UV–Vis diffuse reflectance spectrum (UV–Vis DRS) was
collected using a Shimadzu UV-3100 spectrometer equipped with
an integrating sphere and using BaSO as a reflection standard.
4
1.0 lm) was chosen to separate the mixed components at the fol-
Spectra were recorded in the range of 190–1000 nm in 0.5 nm
steps with a scan speed of 60 nm/min and time constant of 2 s.
In the visible range, a constant slit width of 2 nm was used for
the monochromator, and the reflected radiation was detected with
a photon multiplier detector. The wet pastes were placed on the
window of the integrating sphere for testing.
lowing condition: 423 K for capillary column, 473 K for injector
port, and 523 K for FID detector.
An Agilent 1260 infinity HPLC system coupled with an auto-
sampler, quaternary pump, and ultraviolent detector was used to
determine the concentration of adipic acid and glutaric acid in
the study. The HPLC was performed on a COSMOSIL 5C18-PAQ col-
umn (length, 250 mm; inner diameter, 4.6 mm) and the column
was held at 298 K. The mobile phase was a combination of solvent
A (20 mM potassium phosphate dibasic in water adjusted pH to 3.0
with phosphoric acid) and solvent B (methanol). The gradient anal-
ysis was performed using 84 % solvent A and 16 % B solvent. The
2
.2.6 Combined thermogravimetric (TG) and differential scan-
ning calorimeter (DSC) analyzes were performed on a Mettler-
Toledo TGA/DSC1 instrument at a heating rate of 10 K/min from
room temperature to 1173 K. Its temperature precision was
±0.5 K and microbalance sensitivity was less than ±0.1 lg. Blank
experiments without samples were carried out to obtain the base-
lines to calibrate the experiments with samples at different heating
rates. Under an inner atmosphere of high purity nitrogen (a flow
injection volume was 20 lL. For all gradient segments, the elution
flow rate was 1.2 mL/min, and the detection wavelength was set at
510 nm.