more recently, as potential candidates to be used in the
8
construction of a liquid-mirror lunar telescope.
Q50 thermogravimetric analyser. The measurements were
ꢀ1
done in platinum pans, at a heating rate of 10 K min , in
A series of 1-alkyl-3-methylimidazolium alkanesulfonate
air. The onset of the weight loss in each thermogram was used
as a measure of the decomposition temperature.
ionic liquids, [C mim][C SO ] (n = 1–6, 8, or 10; k = 1–4,
n
k
3
or 6) was prepared (see Fig. 2), and their melting points,
decomposition temperatures, and density and viscosity as a
function of temperature were determined.
Density measurements. Density was measured with an
Anton Paar vibrating tube densimeter, model DMA 5000,
operating at atmospheric pressure and within the temperature
range 293 to 333 K.
Experimental
The internal calibration of the instrument was confirmed by
measuring the densities of atmospheric air and doubly distilled
water, according to the recommendations of the manufacturer.
The DMA 5000 cell was embedded in a metallic block, the
temperature of which was controlled by several Peltier units.
This arrangement allowed a temperature stability better
than ꢂ2 mK.
Instrumental procedures
NMR spectroscopy. All NMR spectra were recorded on a
Bruker Avance spectrometer DPX 300 at 27 1C, using
deuterated trichloromethane as solvent (10–20 mM solutions
1
for H NMR spectroscopy).
Mass spectrometry. All mass spectra of ionic liquids (diluted
solution in ethanenitrile) were recorded using a Waters LCT
PREMIER Electrospray mass spectrometer.
All ionic liquids used in the density determinations were
degassed under vacuum and moderate temperature conditions
for periods longer than 24 h and stored in sealed vials prior to
injection in the densimeter, using non-lubricated disposable
syringes.
Karl Fischer analysis. The water content of ionic liquids was
measured by coulometric Karl Fischer (Metrohm) titration.
Ionic liquids were tested immediately after drying under high
vacuum, and the water content was below 150 ppm for all
samples.
Viscosity measurements. The viscosity was determined using
a rheometer, Rheometrics Scientific SR200, that allows
measurements from 293 to 393 K at atmospheric pressure,
and in a wide viscosity range (1 to 3000 mPa s) depending on
the geometry used. A Couette geometry (concentric cylinders)
was chosen for this study. Temperature was maintained
constant to within 0.01 K by means of a recirculating bath
and was measured with the same accuracy using a platinum
resistance thermometer (PRT) from Hart Scientific (model
Differential scanning calorimetry (DSC). The melting points
of the synthesised ionic liquids were measured by DSC, using
a TA Instruments Modulated DSC 2920. Cooling was
accomplished by using a refrigerated cooling system capable
of controlling the temperature down to 220 K. Dry nitrogen
3
ꢀ1
gas (flow rate of ca. 20 cm min ) was purged through the
DSC cell. The heating/cooling regime was adjusted for each
sample separately.
5
612, accuracy of ꢂ0.018 1C at 0 1C), calibrated against a
secondary reference temperature standard. To decrease the
water contamination of the sample during the measurements,
the rheometer was placed inside a glove-box in an isolating
atmosphere of purified and dried air.
Typically, for solid 1-alkyl-3-methylimidazolium alkane-
ꢀ1
sulfonates, a standard heating and cooling ramp of 5 K min
was used. For liquid samples, including supercooled ionic
liquids, much lower cooling and heating rates were used
In order to improve the accuracy of the viscosity measurements,
the rheometer was calibrated, as a function of temperature,
with the standard viscosity oil N100 from Cannon Instrument
Company (280 mPa s at 293 K).
ꢀ1
(
0.2 to 1 K min ). Both the onset and the peak melting
temperatures were measured and, for each ionic liquid, the
enthalpy of fusion was determined by integration of the
experimental heat flow curves as functions of temperature.
In some cases it was impossible to detect the phase transitions
Viscosity measurements were performed in steps of approxi-
mately 10 K from 293 K up to 360 K in all the ionic
3
liquid samples in which a sufficient amount (ca. 14 cm ) was
(
crystallisation and melting) using the DSC cell. Therefore, for
[
[
C mim][C SO ], [C mim][C SO ], [C mim][C SO ], and
3
2
2
3
2
3
3
2
4
available. Similar measurements performed before in the same
9
equipment permitted a statistical analysis of the results that
C mim][C SO ], seeds were used to trigger crystallisation in
1
6
3
supercooled ionic liquid samples. This procedure could not be
implemented in the DSC cell due to the small size of the
samples used in that case. As a consequence, the samples were
heated slowly in a small glass vial, and melting points were
determined visually as the temperature of the disappearance of
the last crystal. In this case, the uncertainty of the melting
point temperatures (ꢂ3 K in most cases) is much greater than
that of measurements in the DSC cell (or even that of
commercial capillary melting point instruments). The use of
small vials instead of capillary tubes in this case is related to
the handling of the viscous ionic liquid samples.
has indicated a precision in the viscosities of 0.2% and an
overall uncertainty lower than 1%. The same precision cannot
be claimed in the present measurements due to the problems
related to the hygroscopic behaviour of ionic liquids (see two
paragraphs above). An overall uncertainty of around 10% was
estimated in this case, inferred from experiments where the
determinations were cycled over time and temperature in order
to check the reproducibility of the viscosity data.
Materials
The nineteen ionic liquids used in this study were synthesised
following the procedure described below. These include
Thermogravimetric
analysis
(TGA).
Decomposition
temperature measurements were performed in a TA Instruments
[C mim][C SO ],
1
n
=
1–6,
8 or 10; [C mim][C SO ],
n 2 3
n
3
8
940 | Phys. Chem. Chem. Phys., 2009, 11, 8939–8948
This journal is ꢁc the Owner Societies 2009