Full Papers
doi.org/10.1002/cctc.202100309
ChemCatChem
[46]
C D , and CDCl were purchased from Aldrich. Dichloromethane
Wagner’s energy dependence of attenuation length
and a
6
6
3
was distilled over CaH , methanol over Mg, while n-hexane and n-
pentane were used as received. Ketjenblack (C ) was purchased
from Cabot Corp. USA.
standard set of VG Escalab sensitivity factors. The uncertainty on
the atomic quantitative analysis is about �10%. Assignment of
photoelectron signals and peak components was determined
2
K
[47]
according to literature reference database.
1
13
1
H and C{ H} NMR spectra were acquired in C D at room
6
6
Analytical methods
1
temperature by using a Bruker Avance 400 MHz spectrometer. H
NMR spectra were acquired at 400.12 MHz and C{ H} NMR spectra
at 100.0 MHz.
1
3
1
STEM (scanning transmission electron microscopy) images were
acquired with a TESCAN GAIA 3 FIB/SEM apparatus using STEM
detector operating at an accelerating voltage of 30 kV. TEM
analyses were carried out on a TEM PHILIPS CM 12 instrument,
equipped with an OLYPMPUS Megaview G2 camera, and using an
accelerating voltage of 100 keV. Samples suitable for the analyses
were prepared by depositing a drop of a n-pentane solution of 1 or
IR spectra were acquired as KBr pellets at room temperature in the
wave number interval 4000–400 cm using a Perkin Elmer instru-
ment.
À 1
ICP-OES analysis were carried out with an ICP-Optical emission dual
view Perkin Elmer OPTIMA 8000 apparatus.
K
water suspension of 1/Pd@C on a holey film of a 300 mesh Cu
grid, followed by evaporation of the solvent at room temperature.
HPLC analyses were carried out with a Shimadzu apparatus
equipped with a RID 10 A detector and an Aminex HPX-87H column
The mean particle diameter (d ) was calculated by using Equa-
m
tion (1)
(
300 mm×7.8 mm, BIO RAD) using H SO (0.005 M) as eluent with a
2 4
flow rate of 0.4 mL/min and an oven temperature of 35°C.
dm ¼ Sd n =Sn
(1)
i
i
i
GC analyses were carried out with a GC-2010 apparatus from
Shimadzu, equipped with an FID and a capillary column of the type
VF WAXms (30.0 m×0.25 mm×0.25 μm) from Agilent and He as
carrier gas.
with n corresponding to the number of particles with diameter d.
i
i
The total number of particles (N) used for the histogram is reported
in the caption of the corresponding figure.
Powder X-ray diffraction (PXRD) spectra were acquired at room
temperature with a PANalytical X’PERT PRO powder diffractometer,
using CuKα radiation (λ=1.5418 Å) and a parabolic MPD-mirror
and a PIXcel RTMS detector. The spectra were recorded in the 2Θ
range from 5.0 to 80.0° with a step size of 0.105° and a counting
GC-MS analyses were carried out with a QP2010SE apparatus from
Shimadzu, using the same capillary column as for GC analysis.
K
Syntheses of 1and 1@C
time of 428.9 s, using a Si zero background as sample holder.
K
Detailed protocols for the syntheses of L, H L and Pd@C are
4
reported in the Supporting Information (SI).
X-ray photoelectron spectroscopy (XPS)-analyses of the samples
was performed by using a XPS VG MicrotechESCA3000 Multilab
spectrometer, equipped with a twin anode X-Ray source (Mg and
Al) and a five channeltrons detection system. Photoemission
spectra were acquired by using a standard non-monochromatized
Al Kα (hν=1486.6 eV) excitation source, an ultrahigh vacuum
L
0
(343 mg, 0.680 mmol) (Scheme 1) and Pd(OAc)2 (152.6 mg,
.680 mmol) were added to a Teflon-lined autoclave, which was
sealed and evacuated. Afterwards dichloromethane (50.0 mL) was
deaerated and introduced into the autoclave by suction. The
autoclave was then successively stirred at room temperature for
one hour, followed by its pressurization with hydrogen (20 bar) and
mechanical stirring at 35°C for 18 hours. The autoclave was then
successively cooled to 10°C by means of a water-ice bath, the gas
À 6
chamber (base pressure lower than 1×10 Pa) and a hemispherical
analyser operating in CAE mode. The binding energy (BE) scale was
calibrated with Au 4f7/2 core level at 83.9 eV. C 1s peak from the
adventitious carbon (BE=285.1 eV) was used for charge correction
of all spectra. The accuracy of the energy measure is �0.1 eV.
Photoemission data were collected and processed by VGX900 and
XPSPEAK4.1 software, respectively. XPS data analysis was performed
by a nonlinear least square curve-fitting procedure in order to
extract chemical information from the spectral data (i.e., to separate
the photoemission signal originating from distinct elemental or
chemical states). A curve fitting model defined in terms of
component peaks, as a properly weighted sum of Lorentzian and
Gaussian component for the shape of Voight curves, and a Shirley
pressure released, and the dark brown suspension transferred into
a round bottom flask. The solvent was removed by vacuum and the
brown residue suspended in n-pentane (50.0 mL), followed by
filtration over celite. The obtained dark brown solution was
extracted with water (3×10 mL) (i.e., to remove acetic acid, which
was formed during the reaction), dried over Na SO , filtered and
2
4
concentrated to dryness. In order to obtain the product as powder,
the solid was dissolved in dichloromethane (10.0 ml) and then
concentrate again to dryness, obtaining a dark brown solid. Yield:
1
1
10 mg. ICP-OES analysis of 1 showed a Pd content of 23.6 wt%. H
[
45]
background subtraction according to Sherwood has been used.
The fitting parameters that were allowed to vary for a component
peak included the energy position, the full width at half maximum
NMR (C D ) and IR (KBr) spectra are shown in SI, Figure S8 and S9,
6
6
respectively.
K
K
(
FWHM), the percentage of Lorentzian/Gaussian lineshape and area.
1@C was synthesized by suspending Ketjenblack (C ) (1.0 g) in n-
pentane (180.0 mL) followed by sonication for 5 min. at room
temperature, addition of 1 (54.2 mg) dissolved in n-pentane
(30.0 mL), stirring for an hour under a nitrogen atmosphere and
then evaporating the solvent by vacuum. The obtained black solid
was continued drying at 50°C under vacuum. The Pd content of
The software used fitted the experimental data by adjusting all the
parameters of the component peaks until a χ minimum was
2
obtained. The BE position of a component peak resulting from the
curve fitting provided evidence for assignment to an elemental or
chemical environment. Quantitative information concerning the
concentration of the chemical states was inferred by measuring the
relative area of each component peak. C 1s, N 1s and O 1s spectra
showed a single XPS peak, whereas Pd 3d showed a double peak
with well separated Pd 3d5/2 and Pd 3d3/2 spin-orbit components
the isolated solid was analyzed by ICP-OES, giving a Pd content of
1.28 wt%.
(
energy split of ~5.4 eV). Surface relative atomic concentrations
were calculated by a standard quantification routine, including
ChemCatChem 2021, 13, 1–12
9
© 2021 Wiley-VCH GmbH
��
These are not the final page numbers!