The well-known decomposition of tetrahydrofuran (6) by
n-BuLi at room temperature is the third example (Figure 3).8
diagnostic resonance at -0.95 ppm for the RCH2Li protons,
the broad resonances at 3.5 and -0.9 ppm are attributed to
n-BuOLi or n-BuO2Li and its mixed aggregate with n-BuLi,9
respectively. Spectrum (b) reports that 5 min after addition
of THF (6, 4.5 equiv), the onset of its decomposition into
ethylene (7) and the lithio-acetaldehyde enolate (8) is
observable. The RCH2Li resonance is shifted upfield by 0.2
ppm upon addition of THF in response to dynamic THF/n-
BuLi complexation.10 Over time [spectra (c) and (d) at 2
and 48 h] resonances for each of the decomposition products
grew. Even at partial conversion [∼40% in (c)] the large
adjacent THF resonance does not obscure the (O-Z)-vinyl
proton (Hb) of 8. After 48 h all n-BuLi was consumed. Even
though it was not the intent of this ambient temperature
experiment, we subsequently judged the half-life for decom-
position to be ∼2 h, quite consistent with the known value.8d
This well-resolved spectrum of LiOCHdCH2 (8) maintained
its character for at least a week for a sample (again) protected
only with a standard plastic cap.
To further demonstrate the sensitivity and generality of
monitoring reaction progress in nondeuterated solvents, the
fourth example includes spectra of a ring-closing metathesis
reaction (Figure 4). A solution of allyl acrylate (9, 0.5 M in
CH2Cl2) was exposed to 10 mol % Grubbs first generation
ruthenium precatalyst (G1). At 1 min [spectrum (a)],
conversion of a portion of G1 (δ 20.0 ppm) into the Ru-
loaded substrate (13) is evident (δ 18.9 ppm, t, J ) 4.4 Hz).11
Spectrum (b), taken at 4.5 min, shows that G1 is nearly
consumed. The extent of formation of styrene (10) [δ 6.7
(dd, J ) 16.9 and 10.9 Hz) and 5.7 (d, J ) 16.8 Hz) ppm]
correlates with consumption of G1.
Figure 3. Generation (and stability) of ethylene and acetaldehyde
enolate by THF decomposition with n-BuLi.
Formation of γ-butenolide (11) [δ 7.6 (dt, J ) 5.8 and
1.5 Hz) and 6.1 (d, J ) 5.8 Hz) ppm] and E- and Z-dimer
(12)12 is evident and continues as long as resonances due to
any resting carbene are observable. At 64 min [spectrum (c)],
the methylidene carbene 14 (δ 18.9 ppm, s) is the only
identifiable ruthenium-containing species. After 3 h no
carbene resonances remained observable (not shown). At 48
h no further conversion (compared to 3 h) had occurred, so
an additional 10 mol % of G1 was added. One minute later
[spectrum (d)] it was obvious that reaction had been
reinitiated. A similar profile of conversion and alkylidene
Spectrum (a) is of a sample of n-BuLi in hexanes that had
stood in a capped NMR tube for 24 h. In addition to the
(7) (a) Gilman, H.; Gaj, B. J. J. Org. Chem. 1957, 22, 1165-1168. (b)
Honeycutt, S. C. J. Organomet. Chem. 1971, 29, 1-5. (c) Jung, M. E.;
Lyster, M. A. Tetrahedron Lett. 1977, 43, 3791-3794. (d) Stanetty, P.;
Mihovilovic, M. D. J. Org. Chem. 1997, 62, 1514-1515 (t1/2 ) 1.78 h at
20 °C).
(8) There are several methods that can be used to shim the proton probe
to a No-D sample and tube. These include (a) the initial use of a reference
tube of the same volume of deuterated solvent, which is locked and shimmed
as usual and then replaced with the No-D sample; (b) recalling/reentering
a set of previously used shim settings known to be optimal for the solvent
in use; (c) the use of a capillary insert containing a deuterated sample of
the same solvent, which can be locked and shimmed; (d) gradient shimming;
(e) shimming the No-D sample using the FID (see “shimming using the
FID” below); and (f) shimming the No-D sample using the spectrum (see
“shimming using the spectrum” below). Method (c) is a reliable way to
start, and methods (c) and (d) are compatible with autosamplers. We use
methods (e) or (f) for nearly all samples (including all spectra shown here).
Although the following instructions refer to software (VNMR 6.1) on a
Varian, Inc. instrument,5 they are sufficiently generic to be helpful to users
of other manufacturers’ equipment. Initial Setup. Use a typical volume
(e.g., ca. 600-700 µL) of a solution of solute in nondeuterated solvent
with a concentration of, say, 0.1-1 M (ca. 100:1 to 10:1 solvent/solute).
Set the spectrometer parameters as you would for a deuterated sample [e.g.,
nucleus, solvent (if “known” to the spectrometer, otherwise choose
anything), initial shim parameters, spectral width, etc.]. Turn off the “Lock”
but spin the sample. Acquire a single scan spectrum. Phase this initial “un-
easier observation of small changes in the FID level). Adjust the shims,
allowing the FID level to stabilize before making each additional adjustment,
until a maximum FID level (numerical) has been achieved. (f) Shimming
using the Spectrum. Shim by monitoring the increase in reporter peak
intensity (numerical and/or graphical) and peak symmetry. Don’t be
discouraged by initial poor-looking peak shape. In general, note that the
response to a change in shim settings will occur more slowly here than it
will when shimming on a deuterated sample in locked mode. The process
of shim optimization is otherwise quite analogous for samples in nondeu-
terated vs deuterated solvents. Once the shims are optimized, exit “acqi”
and take another one pulse spectrum. If the overall spectrum quality is
acceptable, proceed with recording the spectrum. Shimming the second
through the nth sample at the same sitting will usually be easier.
(9) (a) McGarrity, J. F.; Ogle, C. A. J. Am. Chem. Soc. 1985, 107, 1805-
1810. (b) McGarrity, J. F.; Ogle, C. A.; Brich, Z.; Loosli, H.-R. J. Am.
Chem. Soc. 1985, 107, 1810-1815.
1
shimmed” No-D H NMR spectrum. Select, expand, and note a “reporter
resonance” of known peak shape (a solvent peak is usually a good choice,
but any peak of known multiplicity will suffice). Run the FID/Spectrum
macro gf and enter the interactive acquisition display process acqi (to
allow observation of the real-time FID or spectrum). (e) Shimming using
the FID. After performing the initial setup, select the FID button and then
increase the gain until the FID level is between 500 and 1000 (to allow
(10) Keresztes, I.; Williard, P. G. J. Am. Chem. Soc. 2000, 122, 10228-
10229 and references therein.
(11) Schwab, P.; Grubbs, R. H.; Ziller, J. W. J. Am. Chem. Soc. 1996,
118, 8, 100-110.
(12) Sueltemeyer, J.; Doetz, K. H.; Hupfer, H.; Nieger, M. J. Organomet.
Chem. 2000, 606, 26-36.
Org. Lett., Vol. 6, No. 6, 2004
955