Welcome to LookChem.com Sign In | Join Free

Science Details

Home > Chemical Encyclopedia > Science List > Details
  • Improved neutron–gammadiscrimination at low-light output events using conical trans-Stilbene (cas 103-30-0)

  • Add time:08/13/2019    Source:sciencedirect.com

    Scintillator geometry affects light-collection efficiency (LCE) and, as a consequence, detector-performance metrics that depend on LCE, such as particle-type identification. This work compares the neutron–gammapulse-shape discrimination (PSD) performance of two trans-stilbene crystals of different geometries: a cylinder and cone. Both geometries have a base diameter and height of 50 mm. Results demonstrate that the cone achieved a higher figure of merit over the cylinder by approximately 22.6% for pulses ranging from 25 keVee to 100 keVee. This improvement can be especially helpful for applied detection systems to better resolve neutrons and gamma rays at low-light output events (<100 keVee), where PSD can suffer.

    We also recommend Trading Suppliers and Manufacturers of trans-Stilbene (cas 103-30-0). Pls Click Website Link as below: cas 103-30-0 suppliers

    Prev:Solution growth of a deuterated trans-Stilbene (cas 103-30-0) crystal for fast neutron detection
    Next:Gas-solvent and water-solvent partition of trans-Stilbene (cas 103-30-0) at 298 K)