3
78
P.A.C. McPherson, B.T. Türemen / Biochemical and Biophysical Research Communications 452 (2014) 376–381
2
.4. HPLC of derivatized amino acids
(non-polar) amino acids. Results (Fig. 1) are in keeping with this
prediction, with 6-aminoquinolyl-tyrosine and - -DOPA eluting
mid-profile with average retention times (R ) of 7.8 ± 0.02 and
L
Derivatized amino acids (20
lL sample injection) were resolved
t
on a Phenomenex C18 column (3.9 ꢀ 300 mm; 5
l
m) using a Per-
7.5 ± 0.09 min, respectively (n = 12). These retention times
matched those obtained for a simple sample containing only
kin-Elmer Series 200 HPLC system. A gradient of solvent A (Waters
AccQ.Tag™ Eluent A) in solvent B (60% acetonitrile; Ridel-de Haen,
Steinheim, Germany) was used at a flow rate of 1 mL/min. The
gradient was programmed as follows: 0–5 min, 0–6% B;
L-DOPA and L-tyrosine (Fig. 1, inset). The resolution of the amino
acid derivatives was virtually identical for both detection modes
(UV, Method A and fluorescence, Method B), as expected. The large
5
3
1
–6.5 min, 6–10% B; 6.5–11 min, 10–33% B; 11–13 min, isocratic
3% B. A wash-out sequence of 13–14 min, 33–100% B; 14–
5 min, isocratic 100% B; 15–16 min, 100–0% B completed a sam-
initial peak (R
t
ꢂ 2 min) was the co-product of the derivatization
reaction, N-hydroxysuccinimide, though it appeared that neither
the presence of this additional peak nor those of any of the twenty
common amino acids interfered with the identification of the tar-
get analyte.
ple run. The eluent was monitored by either diode array detection
at kmax 254 nm (Method A) or fluorescence (E 395 nm; E 348 nm)
Method B). For comparison, HPLC employing detection by intrinsic
fluorescence of tyrosine/ -DOPA (E 280 nm; E 320 nm) (Method
C) was also employed as previously reported [16].
x
m
(
L
x
m
3.3. Analytical performance
The measurement of
over the calibration range employed for both Method A and
Method (0.005–100 nmol/L; correlation coefficient of
r = 0.9998) with an identical relative standard deviation (RSD) of
.7% for both methods. This made the routine determination of
-DOPA straightforward as a non-weighted linear calibration func-
tion (peak response vs. concentration) could be used. The limits of
detection (LOD) and limits of quantification (LOQ) were calculated
from the standard deviation of blank samples and the gradient of
the calibration line, assuming a signal-to-noise ratio of 3 and 10,
respectively (Table 1). The repeatability and reproducibility of
the assay, taken as intra-day and inter-day RSD, was similar for
both methods (Table 1). Overall, the performance characteristics
for Methods A and B were superior to those found for the assay
employing intrinsic fluorescence (Method C), particularly with
regard to the LOQ and inter-day CV.
In establishing our assay, we selected detection by either UV or
fluorescence as at least one of these detection modes will be avail-
able on even the most basic HPLC instrument and is therefore
accessible to a wide-range of laboratories. We did not investigate
the use of electrochemical detection as this is generally compara-
ble to fluorescence detection in terms of sensitivity, as shown by
Pappa-Louisi et al. [32] and would therefore offer no analytical
advantage. We found that the anticipated detection method had
an impact on the derivatization procedure employed, as prelimin-
ary experiments utilizing neat AccQ reagent and UV detection
demonstrated poorer resolution of peaks, due to the dominance
of the N-hydroxysuccinimide peak. It was found that when using
UV detection, a 1/10 dilution of AccQ reagent could be used with
no observable effect on the derivatization reaction, yet substan-
tially improving the resolution of peaks. The neat reagent was
required for optimum detection by fluorescence down to the
pmol/sample level.
L-DOPA was found to be completely linear
2
.5. Miscellaneous methods
B
Independent measures of protein oxidation were used at vari-
2
L
ous stages in the development of the analytical procedure. Total
protein carbonyls were determined spectrophotometrically as
their 2,4-dinitrophenylhydrazone derivatives [28]. The concentra-
tion of
L-tyrosine and L-DOPA were determined by commercially
available enzyme-linked immunosorbent assays (tyrosine:
Immundiagnostik AG, K7015;
MBS162177).
L-DOPA: BioSource ELISA kit,
2.6. Method validation and statistical analysis
Routine descriptive statistics were generated from analysis of
data using Microsoft Excel (2007) and results expressed as
mean ± standard deviation. Non-parametrically distributed results
were analyzed by the Mann Whitney U-test. A comparison of the
methods was achieved by two-way analysis of variance (ANOVA).
Both statistical tests were performed using the Statistics Package
for Social Sciences (SPSS) Version 17.0 for Windows. P < 0.05 was
considered as statistically significant.
3
. Results and discussion
3.1. Preparation of BSA sample material
To provide readily available source of sample material for the
development and characterization of the assay, BSA was oxidized
by alkylperoxyl radicals generated by the thermal decomposition
of the azo-initiator AAPH [29]. The levels of protein carbonyls
were assessed to verify adequate oxidation of the protein, and
provided results in keeping with those available in the literature
(
2
4.9 ± 0.2 nmol/mg BSA) [30]. The levels of
L-tyrosine (30.9 ±
3.4. Impact of cloud-point extraction
.3 nmol/nmol BSA) and -DOPA (39.6 ± 2.3 pmol/nmol BSA) were
L
determined by ELISA and provided an estimate of the expected
concentrations of these analytes.
When dealing with samples of a more complex nature (BSA or
human plasma), it was found that detection of tyrosine and L-DOPA
was limited to the submicromolar range for UV detection and the
nanomolar range for fluorescent detection, and thus did not repre-
sent a particularly significant improvement. However, when sam-
ples were pre-concentrated using cloud-point extraction, the
limit of detection was shifted by an order of magnitude for both
3
.2. Reversed-phase HPLC of derivatized amino acids and
L-DOPA
A set of twenty standard amino acids plus -DOPA (2.5 mol/L)
L
were derivatized and subjected to HPLC initially using conditions
recommended by the manufacturer [31]. The modification of these
conditions reported in the Experimental section afforded good res-
olution of the amino acids peaks, with peak heights in keeping with
the derivatization chemistry. The position of the L-DOPA peak was
predicted to be in advance of tyrosine, based on the presence of an
additional –OH group which increases the polarity of the molecule,
detection methods and enabled detection of
BSA using UV detection (Table 2).
L-DOPA in unoxidised
The effect of cloud-point extraction is twofold. Firstly, as Triton-
X114 is amphiphilic, micelles are formed in aqueous solution with
a highly hydrophobic core. The non-polar components of the reac-
tion mixture (mainly AccQ-derivatives) migrate from the aqueous
buffer solution to the hydrophobic core, separating them from
resulting in L-DOPA eluting before tyrosine and the remaining