5794
A. Amirav et al. / Tetrahedron 68 (2012) 5793e5799
(SMB) (also named Supersonic GCeMS).1e6 Supersonic GCeMS is
based on a GC and MS interface with SMB and on the electron
ionization of vibrationally cold analytes in the SMB (cold EI) in
a fly-through ion source. This ion source is inherently inert and
further characterized by an ultra fast response time and vacuum
background filtration capability.1,7 The same ion source also offers
a mode of classical EI.8 Cold EI, as a main mode, provides trust-
worthy enhanced molecular ions combined with effective library
sample identification, that is, supplemented and complemented by
a powerful isotope abundance analysis method and software,9
which provides elemental formulas with unit resolution quadru-
pole MS data. In addition, the range of low volatility and thermally
labile compounds amenable for analysis is significantly increased
(about doubled) with Supersonic GCeMS due to the use of contact-
free fly-through ion source and the ability to lower sample elution
temperatures through the use of short GC columns with high
carrier gas flow rates.5 This compatibility with high column flow
rates further facilitates effective fast and ultra fast GCeMS ana-
lysis,1e3,10 which was also explored in its use in GCeMS of reaction
products.6
The basic requirement from a system that should measure
reaction yields is that its response will be uniform and
compound independent or the response should represent in
a known and systematic manner the compounds molecular weights
and amounts.
Electron ionization, unlike any other ionization method,
uniquely provides ionization yields that depend predominantly on
the sum of the numbers of each different atom in the ionized
analyte11e13 due to the localized nature of electroneelectron in-
teraction in the ionization process. In simple terms, the electron
ionization cross section approximately depends on the number of
electrons in the molecule, which relates to its molecular weight.
Thus, the electron ionization probability for organic compounds
approximately linearly increases with the ionized compound’s
molecular weight hence it is approximately proportional to the
sample weight. Consequently, it is well known that standard
GCeMS exhibits uniform response to all volatile compounds found
in analyzed samples.
For GCeMS analysis, that is, based on a quadrupole mass
spectrometry, the ionization yield is further reduced by the mass
analyzer and ion detector transmission, hence its response slightly
declines with mass. However, for semi-volatile and low volatility
compounds, peak-tailing and ion source degradation can signifi-
cantly reduce the observed signal, sometimes in a difficult to
predict way, thus the feature of uniform response of standard
GCeMS is eroded and often does not exist anymore. In contrast, in
GCeMS with SMB analysis, ion source-related peak-tailing and
degradation are completely eliminated, due to the use of a fly-
through ion source, thus providing practically uniform com-
pound independent response for all analytes that enter the SMB.
The implication of such uniform response is that in a given chro-
matogram, the peak area relates to the amount (weight) of the
analyte.
Fig. 1. The analysis of a mixture of n-C16H34, methyl stearate, cholesterol and n-C32H66
at about 0.25 ng on-column amount with standard GCeMS (upper trace) and
5975-SMB Supersonic GCeMS (bottom trace). 30 m, 0.25 mm ID capillary column with
0.25
m DB-5ms UI column was used with the standard GCeMS with 1.2 mL/min column
flow rate while only 5 m of that column was used with 5 mL/min column flow rate
with the Supersonic GCeMS. The GC oven program was 80 ꢀC followed by 10 ꢀC/min to
320 ꢀC for the standard 5975 GCeMS and 50 ꢀC followed by 30 ꢀC/min to 300 ꢀC for
the 5975-SMB.
EI mass spectra while for standard EI it was either weak or absent
in the case of n-C32H66
.
Thus, the uniform response of GCeMS with SMB can be as-
sumed and routinely used for semi-quantitative analysis. Further-
more, a simple chromatogram with few compounds at various
molecular weights could be utilized for obtaining more accurate
calibration of the cold EI system response for closer accounting of
small experimental mass dependent effects on the response factors
dependence on mass.
In this work we describe the use of GCeMS with SMB for the
semi-online monitoring of organic reactions (fractions are taken
during the reaction time and progress), while optimizing these
reactions yields, obtaining information on their mechanisms and
determining products purity and identity, including in some cases
information regarding existence and possible structures of formed
isomers.
In Fig. 1, the uniform response of our 5975-SMB Supersonic
GCeMS is demonstrated via the comparison of two chromato-
grams of the same test mixture that included n-C16H34, methyl
stearate, cholesterol and n-C32H66, each at 0.25 ng on-column
amount (n-C32H66 amount is somewhat lower). As clearly ob-
served, the response of standard EI is declining with the sample
volatility and mass, and already for cholesterol its relative re-
sponse has declined by more than an order of magnitude in the
total ion count scale (TIC) (factor of 10.2 in relative peak area). In
contrast, the response of 5975-SMB remained unaffected by the
sample volatility due to the use of contact-free fly-through ion
source and it is very similar to that of GC-FID. In addition, all the
four sample compounds had dominant molecular ion in their cold
2. Experimental: the Supersonic GCeMS systems
2.1. General
Two types of GCeMS with Supersonic Molecular Beams (SMB)
(Supersonic GCeMS) were used. The first system was based on the
conversion of a Varian 3800 GC plus 1200 triple quadrupole MS
(Varian Walnut Creek CA USA) into a Supersonic GCeMS as de-
scribed in details in Ref. 4. The second system was a recently de-
veloped Supersonic GCeMS based on the conversion of an Agilent
7890 GCþ5975 MSD (Agilent Technologies, Santa Clara CA USA)
into
a 5975-SMB Supersonic GCeMS (Aviv Analytical, Hod
Hasharon Israel).