Low-Temperature Pyrolysis of Vinyl Bromide
J. Phys. Chem. A, Vol. 105, No. 10, 2001 1831
kJ mol-1 with a preexponential factor of 9.5 × 1010 L mol-1
s-1 for this reaction.
While the thermal decomposition of C2H3Br has been studied
from approximately 900 to over 2000 K, no evidence has been
reported for a decomposition channel other than the molecular
one leading to C2H2 and HBr. This contrasts with our observa-
tion of substantial yields of C2H4 and 1,3-C4H6 in the decom-
position at temperatures as low as 630 K, suggesting the
occurrence of a free radical decomposition channel that produces
C2H3 radicals. There is also some disagreement concerning the
activation energy for the molecular process. We have therefore
undertaken an experimental study of the lower temperature
decomposition of vinyl bromide below 800 K to measure the
relative importance of the molecular and free radical channels
and to clarify the kinetics and mechanism of the free radical
process. We report here a kinetic analysis of our data which
separates the molecular and free radical channels and provides
an empirical kinetic characterization of each.
Figure 1. Effect of varying the surface/volume ratio. Open symbols
are for the unpacked reaction vessel, S/V ) 1.0 cm-1, at a temperature
of 711 ( 2 K and a pressure of 16.1 ( 0.4 kPa of C2H3Br. The larger
closed symbols are for the packed reaction vessel, S/V ) 11 cm-1, at
a temperature of 704 ( 2 K and a pressure of 16.2 ( 0.6 kPa of C2H3-
Br. The smaller closed symbols are for the packed reaction vessel,
corrected to refer to the temperature of 711 K used in the unpacked
reaction vessel, using the experimental activation energies for each of
the reaction products. O, b, C2H2: 0, 9, C2H4: 4, 2, 1,3-C4H6.
Experimental Section
The pyrolysis experiments were made in Pyrex reaction
vessels that were heated electrically and were well-insulated.
The surface/volume ratios of these reactors ranged from 1.0 to
11 cm-1. The temperature was monitored continuously with
calibrated iron-constantan thermocouples that were inserted into
thermocouple wells extending into the center of the pyrolysis
vessel. The temperature remained constant during an experiment
to within the measurement precision which was better than
(0.5%. Pressures were measured with piezoelectric pressure
transducers whose calibration was checked daily. Kinetic
measurements made under identical conditions in different
reaction vessels were indistinguishable.
Experiments were all made in batch mode. A reaction mixture
of known composition was introduced to the reaction vessel at
a measured pressure. It was then allowed to react for a measured
length of time that was at least 10 times the settling time for
the pressure. At the end of the desired reaction time, a measured
pressure of the reaction mixture was transferred to a gas sample
loop and injected onto the chromatographic column. This
procedure prevented loss of products such as ethylene that do
not trap quantitatively with liquid nitrogen. When reactions
required the use of mixtures of vinyl bromide with other gases
such as argon, the gases were added to a reservoir at known
pressures and were allowed to mix for a sufficiently long time
that experiments using both longer and shorter mixing times
gave identical results.
the reaction products were determined initially by comparison
of their chromatographic retention times with those of authentic
samples, and these identifications were confirmed by obtaining
the mass spectra of the individual product peaks as they eluted
from the chromatographic column. The resolution of the
products was also checked by observing the mass spectra
obtained during successive scans of the mass spectrum as a
given peak eluted from the column.
Vinyl bromide was obtained from the Aldrich Chemical Co.
and had a stated purity of 98%. It contained methyl ethyl
hydroquinone as an inhibitor. It was purified before use by
freeze-pump-thaw cycles followed by bulb to bulb distillation
retaining only the middle part. Gas chromatographic analysis
of the purified material on the SP1700 column gave only a single
peak, while analysis on the Hayesep T column gave a small
additional peak, representing about 0.2% of the C2H3Br peak,
with a retention time slightly less than that of the C2H3Br peak.
This additional peak did not interfere with those of the reaction
products being measured and its area, relative to that of the
C2H3Br peak, did not change when the vinyl bromide was
pyrolized. It was concluded that the substance responsible for
the small additional peak when the Hayesep T column was used
for analysis did not affect either the chemical analysis or the
chemical behavior of the reaction under study.
The reaction mixtures were analyzed by gas chromatography
using either a Perkin-Elmer Sigma 3 or a Hewlett-Packard
HP5880 gas chromatograph. Both instruments used a flame
ionization detector, and the HP5880 was also interfaced to a
Ametek Dycor MA200M quadrupole mass spectrometer. Most
Results and Discussion
1
of the analyses used a 30 foot, /8 in. diameter column packed
Ethylene, acetylene, 1,3-butadiene, and vinyl acetylene were
the hydrocarbon reaction products obtained in the experiments.
Their yields were measured as a function of both the pressure
of vinyl bromide and reaction time at each temperature. The
extent of decomposition of vinyl bromide was at most 5% and
in most cases was much less than this. While HBr is presumably
formed in the reaction, we were not able to detect it or measure
its concentration with the available instrumentation. Representa-
tive plots of product yields as a function of reaction time are
shown in Figures 1 and 2. The standard deviations of the points
on such plots were typically of the order of 10-25%, depending
on the temperature, specific reaction product, and extent of
with SP1700, but a few analyses used columns packed with
Carbosieve G, Porapak Q, or Hayesep T to verify acceptable
resolution of the components to be analyzed in the reaction
mixtures. The chromatographs were calibrated daily with gas
mixtures containing known amounts of the desired analytes
diluted in helium. Typical chromatographic conditions with the
SP1700 column were 75 psi helium carrier, temperature program
50 °C for 20 min, ramp to 70 °C at 10 °C/min, and hold at 70
°C for 30 min. Typical retention times under these conditions
were as follows: ethylene, 6.5 min; acetylene, 8.7 min; 1,3-
butadiene, 20.7 min; vinyl bromide, 46 min. The identities of