Larsen, Haulena, Grindem, Gulland
determine whether they were fit for release. Determin-
laboratory personnel. Laboratory quality control (QC)
was performed, with calibration on every lot of samples
and also every 24 hours. Once per month, blind studies
were conducted, and QC values, methods, and ranges
were reviewed. Blood from the EDTA tube was used to
determine HCT using an automated cell counter (Vet
ABC, Heska Corporation, Fort Collins, Colo, USA). The
cell counter had previously been calibrated for marine
mammal RBCs. Calibration had been established by
comparing cell counter results from juvenile elephant
seals (n=19) with HCT values obtained using microcen-
trifugation. For elephant seals, values from the cell
counter were 0.011 (± 0.011) L/L higher than manually
spun results (J. Lawrence, Marine Mammal Center, per-
sonal communication).
ation of an animal’s suitability for release was based on
clinical assessment, physical examination, and evalua-
tion of CBC and serum chemistry results.
At the time of sampling, the seals weighed 67 ± 6 kg
and were estimated to be 4-5 months old. They were
manually restrained, and blood was drawn from the
epidural intravertebral vein into a collection tube coated
with lithium heparin (Vacutainer, Becton Dickinson,
Franklin Lakes, NJ, USA) using Monoject needles (20 ga
by 3.8 cm) and adapters (Sherwood Medical, St Louis,
Mo, USA). Lithium heparin was chosen over sodium
heparin so electrolyte analysis would not be altered by
the electrolytes in the heparin medium. Samples were
analyzed within 10 minutes with the portable clinical
analyzer (i-STAT, i-STAT Corporation, East Windsor, NJ,
USA) by introducing heparinized whole blood (approx-
imately 65 µL) into a disposable cartridge (i-STAT 6+, i-
STAT) designed for simultaneous assay of electrolytes,
glucose, BUN, and HCT. Each cartridge contains a series
of thin-film electrodes (biosensors) that contact the
blood sample and send signals to an electronic system.
The system compares these signals with calibration sig-
nals contained within the cartridge and processes the
Mean, SD, median, 25th percentile, 75th percentile,
and minimum and maximum values for each blood ana-
lyte using each method of analysis were determined.
The mean (± SD) difference between values obtained for
the same sample (value from the portable analyzer
minus the value from the automated analyzer) were
determined for all analytes.The intervals of values from
the 2 methods were compared with previously reported
4
intervals for juvenile elephant seals.
12
results. Sodium (mmol/L), potassium (mmol/L), and
chloride (mmol/L) values were measured by direct ion-
Results and Discussion
12
selective electrode potentiometry. Glucose was mea-
sured by oxidation with glucose oxidase and ampero-
Sodium values from the portable analyzer were slightly
lower than those previously reported, but values from
1
2
metric measurement of hydrogen peroxide. Urea
BUN) was hydrolyzed to ammonium in a reaction cat-
alyzed by urease, and the resultant ammonium ions
4
(
the automated analyzer were slightly higher (Table 1).
Portable analyzer values were 4-10 mmol/L less than
those from the automated laboratory instrument, sug-
gesting that sodium values obtained with the 2 tech-
niques cannot be considered equal. Equating the 2 mea-
surements could cause a clinician to make an erroneous
diagnosis of pinniped hyponatremia (sodium < 147
11
were measured amperometrically. Values for glucose
and BUN were reported by the analyzer as mg/dL;
results were then converted to mmol/L. HCT (L/L) was
12
determined by conductivity.
At the same time that the heparinized sample was
taken, a second blood sample was collected from each
patient into a plain red-top glass tube (Vacutainer) and
into a tube containing EDTA (Vacutainer). These sam-
ples were kept refrigerated and were processed by
trained laboratory personnel within 1 hour of sample
collection. The samples in the plain tubes were allowed
to clot, centrifuged at 3100g for 10 minutes, and then
serum was extracted using a pipette. Serum concentra-
tions of sodium, potassium, chloride, BUN, and glucose
were measured using an automated analyzer (Olympus
AU5200, Olympus America, Melville, NY, USA). Sodi-
um, potassium, and chloride concentrations were deter-
mined by indirect ion-selective electrode potentiometry.
6
mmol/L) and could lead to inappropriate treatment.
The cause for the discrepancy probably is related to the
use of heparinized whole blood rather than serum in the
portable analyzer. Elephant seal RBCs have relatively
high intracellular sodium concentrations relative to ter-
14
restrial mammals, and some of this sodium is released
when blood samples are spun for serum extraction.
However in a previous study in domestic dogs, there
was poor correlation for sodium values between the i-
STAT and an automated chemistry analyzer, even when
10
heparinized whole blood was used for both machines.
Therefore it is prudent to interpret laboratory values,
especially sodium values, using appropriate specimen
and instrument-derived reference intervals.
13
Glucose was measured using the hexokinase method.
BUN was determined colorimetrically with reductive
amination of 2-oxoglutarate by glutarate dehydrogenase
oxidation using ammonia generated by urea degrada-
tion with urease.The samples were processed by trained
Both techniques produced potassium values com-
parable to previously reported serum potassium values
for elephant seals (Table 1). However, potassium con-
4
centrations from the portable analyzer were 0.45 (± 0.43)
Vol. 31 / No. 3 / 2002
Veterinary Clinical Pathology
Page 107