60 J. Phys. Chem. A, Vol. 104, No. 1, 2000
Hakoda et al.
products. In the present work, gas-phase TCE was irradiated
with EB. The decomposition efficiencies were examined under
various initial TCE concentrations and water concentrations. The
products were analyzed and the collection with an alkaline
solution was investigated. The decomposition mechanism was
discussed on the basis of product distribution.
Experimental Section
All sample gas was prepared by mixing dry air (<1 ppmv
CO, <1 ppmv CO2, and <1 ppmv CH4) and vaporized TCE
(>99.0% pure liquid TCE without further purification, Kanto
Chemical Co. Ltd.). Pyrex glass vessels were used as an
irradiation reactor.8 The reactors (50 × 50 × 200 mm) have a
volume of 500 mL and two stopcocks (without grease) at both
ends. The sample gases were introduced into the reactors and
were sealed with the stopcocks. The concentrations of TCE in
the sample gases were measured by a gas chromatograph (GC-
8A, Shimadzu Seisakusho Co. Ltd.) with FID and a packed
column (BX-20 100/120, 2 mm i.d. × 3 m, GL Science Inc.),
and the values were adjusted to 50-1800 ppmv. The water
concentrations in the dry air were measured to be 200 ppmv by
a moisture analyzer (Hygro-3M with D-2 detector, General
Eastern Co. Ltd.). Vaporized water from a liquid water bubbling
bottle was added to the sample gas, and the concentrations were
adjusted to 400-25 000 ppmv.
The sample gases in the reactors were irradiated with 2 MeV
electrons generated by an electron beam accelerator (3 MV 25
mA max., Cockcroft-Walton type, Radiation Dynamic, Inc.)
with a constant beam current of 1.0 mA. The reactors were put
on the conveyer for irradiation and passed under the scan horn.
The sample gases in the reactors were exposed at doses ranging
from 2.2 to 40 kGy. An average dose in the reactor was
measured by a CTA film dosimeter and was 2.2 kGy/pass under
these conditions. A calculated value of average dose rate was
2.1 kGy/s.
Figure 1. Concentrations of residual TCE and the products versus
dose for EB irradiation of 650 ppmv TCE: (O) TCE; (4) dichloroacetyl
chloride (DCAC); (3) CO; (0) CO2; (]) phosgene (COCl2); (∞) TC
1
(total carbon concentration) × /2.
Ozone (O3) was analyzed by an ozone analyzer (model 4560
Sensor STIK with Type 4501 Monitor, EIT).
Result
Figure 1 shows concentrations of residual TCE and products
versus dose at an initial concentration of 650 ppmv. Products
were identified as dichloroacetyl chloride (CHCl2COCl, DCAC),
carbon monoxide (CO), carbon dioxide (CO2), phosgene
(COCl2), and small amount of chloroform (CHCl3). The concen-
tration of total carbon (TC) remained constant up to 40 kGy so
that any products containing carbon did not attach on the inner
surface of the reactor. The concentration of DCAC increases
and reaches a value of 390 ppmv (62%v of initial TCE
concentration) at doses ranging from 0 to 4.4 kGy where TCE
is decomposed. Then, the concentration gradually decreases at
more than 4.4 kGy. For CO and CO2, the concentrations rapidly
increase up to 2.2 kGy and linearly increase at doses ranging
from 2.2 to 40 kGy. For COCl2, the concentration increases at
doses ranging from 0 to 4.4 kGy and remains constant (ca. 90
ppmv). The concentration of CHCl3 was measured to be 5 ppmv
(0.8% v of initial TCE concentration) at 20 kGy.
After irradiation, a rubber cap was attached to one side of
the connection tubes of the reactor and the air between the rub-
ber cap and the stopcock was evacuated by a syringe. Then,
the stopcock was opened and a syringe sampled 2.0 mL of the
inner gas.
Analyses of residual TCE and products were carried out as
follows.
Figure 2 shows the carbon concentrations for TCE and
products in the gas phase with and without NaOH solution.
Figure 2A (without NaOHaq) corresponds to Figure 1. In this
figure, the values in parentheses denote mass yield. The mass
yield stands for the proportion of summation of the carbons
contained in residual TCE and product molecules to input TC.
These values are close to 100% so that all carbon compounds
could be identified. As shown in Figure 2B (with NaOHaq),
DCAC, CO2, and COCl2 could be dissolved in NaOH solution.
These compounds can be mineralized by hydrolysis and
converted into sodium dichloroacetate (from DCAC) and NaCl
and Na2CO3 (from CO2 and COCl2), respectively. The smallest
mass yield after treatment with NaOHaq occurred at 4.4 kGy,
which is where the largest removal efficiency of total VOCs
including products was obtained. Trichloroethylene was almost
completely decomposed at this dose. At more than 4.4 kGy,
the efficiency gradually decreased with an increase in the
concentration of CO. The solutions were also analyzed by the
ion chromatograph to measure chloride ions (Cl-). The result
is shown in Table 1. The expected concentration denotes the
concentration of Cl- ions produced from DCAC, COCl2, and
HCl and/or Cl2 when these products were dissolved in 10 mL
of NaOH solution.
Trichloroethylene, dichloroacetyl chloride (DCAC), and
chloroform (CHCl3) were analyzed by the gas chromatograph.
Other organic gaseous products were barely observed. Total
carbon (TC) and carbon dioxide (CO2) were analyzed by a total
organic carbon (TOC) analyzer (TOC-5000A, Shimadzu Sei-
sakusho Co. Ltd.). The TOC analyzer can convert all carbon
compounds into CO2 in a heating tube (923 K) with a catalyst.
The resulting CO2 formed in a TOC analyzer corresponds to
TC. Carbon monoxide (CO) was analyzed by a CO analyzer
(TH-D4 detector with TA-470 Monitor, Komyo Co. Ltd.).
Phosgene (COCl2) was indirectly analyzed by a TOC analyzer.
When COCl2 is dissolved in phosphoric acid solution (25% w
H3PO4), CO2 is produced from COCl2 in the solution and
released into the gas phase. The concentration of COCl2 was
calculated from the difference in CO2 concentrations with and
without H3PO4 solution. Hydrochloric acid and/or Cl2 were
analyzed by an ion chromatograph (LC-10A, Shimadzu Sei-
sakusho Co. Ltd.) with a conductivity detector (CDD-6A,
Shimadzu Seisakusho Co. Ltd.) and a column (Shim-Pack IC-
A3, Shimadzu Seisakusho Co. Ltd.). The reactors were washed
with NaOH solution (0.1 M) of 10.0 mL to mainly scavenge
Cl2 and HCl. A part of the solution was analyzed for Cl- ions.