B. Nigovi´c, S.B. Hocevar / Electrochimica Acta 58 (2011) 523–527
525
-6
-4
-2
0
Evidently, the electroanalytical performance of the SbFE for mea-
suring sulfasalazine is advantageous over that observed at the BiFE
and, particularly, over that of bare GCE. This behavior suggests that
the electrochemical reduction of the examined drug molecule at
the selected electrodes, under identical conditions, must be the
same process, however, the electron transfer kinetics is obviously
higher in the case of SbFE. Furthermore, preliminary CV studies
revealed the improved electroanalytical performance of the SbFE
due to its favorable surface stability providing great possibilities for
convenient repetitive voltammetric measurements without time-
consuming surface cleaning step as in the case of GCE (inset of
Fig. 2). Namely, strong adsorption of sulfasalazine onto the sur-
face of bare GCE induced significant decrease of the corresponding
voltammetric signal after each successive scan.
1
2
-12
-8
-4
0
-0.2
-0.6 -0.8 -1.0
E vs-0(A.4g/AgCl)/V
3
-0.2
-0.4
-0.6
-0.8
-1.0
E vs (Ag/AgCl)/V
Fig. 2. Comparison of voltammetric signals for 5 × 10−4 M sulfasalazine obtained
at the antimony film (1), bismuth film (2) and bare glassy carbon electrode (3) in
Britton–Robinson buffer solution with pH 4.0. SWV settings are as in Fig. 1. Inset
depicts successive cyclic voltammograms of 5 × 10−4 M sulfasalazine recorded at the
2 × 10−4 M sulfasalazine on the parameters such as concentra-
tion of the antimony ions in the plating solution, deposition time
and deposition potential associated with the antimony film prepa-
ration (Fig. 3). These experiments were conducted to find the
optimum conditions for the antimony film deposition via study-
ing the reproducibility of cathodic current response of sulfasalazine
for ten successive measurements for each prepared electrode. On
the basis of this study, the optimal concentration of antimony ions
for the ex situ film deposition was 20 mg L−1. The voltammetric
signal increased up to 40 mg L−1 of antimony ions in the plat-
ing solution (Fig. 3A), however, lower concentration of antimony
resulted in a more stable film. The effect of deposition time was
also investigated as another important parameter influencing the
quality of antimony coating. As demonstrated in Fig. 3B, the sig-
nal increased up to 90 s of deposition time, however, at longer
deposition times the reproducibility of current response decreased,
possibly as a consequence of thicker antimony film which, as antic-
ipated, became less stable. Finally, the deposition potential was
examined in the potential window of −0.5 V to −1.1 V in combina-
tion with the deposition time of 60 s (Fig. 3C). Evidently, the SWV
signal increased while decreasing the deposition potential from
−0.5 V to −0.7 V, whereas at more negative deposition potentials
the signal attenuated. Furthermore, the current response started to
decrease significantly after applying more negative potentials than
−0.9 V. The antimony film with optimal properties was attained
using the deposition potential of −0.7 V for 60 s in the plating
solution containing 20 mg L−1 antimony(III). The SbFE prepared
ex situ according to these optimized conditions exhibited favor-
able electroanalytical performance for measuring drug molecule
with a well-developed signal and an excellent reproducibility. The
relative standard deviation (RSD) of the peak current was 0.7%
(mean ip = 6.18 A) and the RSD of corresponding peak potential
was found to be 0.5% (mean Ep = −0.475 V). The stability of SbFE
was further evaluated by monitoring a voltammetric response dur-
ing increased number of consecutive measurements at the same
antimony film. These recordings indicated that the signal was fairly
constant up to 40 repetitions in combination with 10 s stirring step
between each measurement with calculated RSD of the peak cur-
rent of 0.9%. The signals started to decrease slightly thereafter
indicating the need for re-plating a new antimony film onto the
supporting electrode surface.
SbFE in Britton–Robinson buffer solution (pH 5.0) using a scan rate of 100 mV s−1
.
then gradual decrease with increasing pH value due to involvement
of a proton transfer in the examined reduction process. Considering
the pH dependence of current response for sufasalazine at the SbFE,
a Britton–Robinson buffer solution with pH 4 was chosen for further
obtained was 59.4 mV pH−1 indicating a 2e−/2H+ electro-reduction
process, which can be attributed, in accordance with previous stud-
ies, to the reduction of an azo-bond of the examined drug molecule
to corresponding hydrazo-species (Scheme 1) [12]. Also successive
cyclic voltammograms (CV) showed one well-defined irreversible
reduction peak, as depicted in the inset of Fig. 2. Undistorted signals
obtained with subsequent potential scans suggested the possibility
of achieving highly reproducible response at the antimony film. CV
behavior was examined further as a function of a scan rate in order
to obtain additional insights into the redox reaction under inves-
tigation. Variation of the scan rate in the range of 20–400 mV s−1
resulted in a linear relationship between the cathodic peak current
and the square-root of a scan rate. Additionally, the peak potential
was shifted towards more negative potential values with increasing
the scan rate. These observations imply on a diffusion-controlled
irreversible reduction process at the SbFE.
To improve the sensitivity, SWV was employed for measur-
ing sulfasalazine due to the most favorable response compared to
other voltammetric modes such as differential pulse voltamme-
try and linear-sweep voltammetry. Variation of the peak current
for sulfasalazine recorded at the SbFE was monitored while chang-
ing instrumental parameters, i.e. frequency, pulse amplitude and
potential step. The peak current was observed to increase linearly
upon raising the frequency in the range of 20–120 Hz, and at fre-
quencies higher than 100 Hz, the signal was obscured by a large
background current. When the pulse amplitude was varied in the
range of 10–90 mV the voltammetric signal was optimal for the
amplitude of 50 mV. The influence of potential step was investi-
gated in the interval from 1 to 6 mV. Finally, the most favorable
100 Hz, a pulse amplitude of 50 mV and a potential step of 2 mV.
A comparative study of the SbFE, BiFE and bare glassy car-
bon electrode for measuring 5 × 10−4 M sulfasalazine as a model
organic molecule is presented in Fig. 2. The newly introduced SbFE
exhibited a very similar behavior to that of the BiFE considering
position of the cathodic voltammetric signal, however the SbFE
provided approximately 35% higher current response compared to
the signal obtained at the BiFE. In comparison with bare GCE, the
peak potential of sulfasalazine obtained at the SbFE was ca. 50 mV
less negative and the voltammetric signal was higher for ca. 40%.
of the SbFE was monitored with respect to the increas-
ing concentration levels of the analyte. An excellent linear
response in the concentration range of 3 × 10−6–2.5 × 10−4 M was
obtained, as shown in the inset of Fig. 4. The calibration plot
is described by the following regression/calibration curve: ipa
(A) = 2.16 × 104C − 0.133 (r = 0.9991), whereas the standard devi-
ations for the slope and the intercept of the calibration curve
were 3.99 × 102 and 0.0056, respectively. For comparison, the BiFE