Reaction of NO2 on Hexane Soot
J. Phys. Chem. A, Vol. 104, No. 51, 2000 11927
TABLE 1: Knudsen Cell Reactor Parameters
is O-ring sealed by a blank flange. Table 1 lists the Knudsen
cell parameters used in this study. The most important param-
eters for the Knudsen cell are the escape constant, kesc, and the
residence time of the molecules inside the reactor, τ, where τ
) 1/kesc.
Reactor parameter
Value
Volume, V
Total calculated surface area
964 cm3
1002 cm
2
2
Geometric area of the sample holder, A
Effective area of the escape aperture, Ah,eff
s
11.95 cm
a
2
0.0677 cm
When the sample compartment is opened to a steady-state
flow of NO2, loss of NO2 to soot surface competes with escape
through the exit aperture, causing a decrease in the mass
spectrometer signal. Thus, the geometric uptake coefficient, γg,
also referred to as the observed uptake coefficient, γobs, is
0.651 s (τ ) 1.5)c
-
1 b
k
esc for NO
2
.370 s (τ ) 2.7)c
-1 d
0
a
b
Effective area after accounting for the Claussing factor.
k
esc
)
1/2
c
d
cj Ah,eff/4V; cj )(8RT/πM) . τ ) 1/kesc
. kesc experimentally determined.
2
0,21
determined from eq 1
geometric area of soot was used in calculating γ. This may
explain the very high γ values reported in some studies
compared to others. In addition, in all of the studies discussed
above, surface areas were not measured for the soot samples
under investigation. This is usually done using the BET method
Ah,eff Io - I
γg )
) γobs
(1)
(
)
As
I
1
8
where Ah,eff and As are listed in Table 1, Io is the mass
spectrometer signal for NO2 prior to opening the sample
compartment, and I is the mass spectrometer signal when the
soot is exposed.
and N2. Instead, values were taken from the literature, which
may or may not accurately represent the true values because of
differences in sample preparation. The fractal nature of soot
also makes the characterization of the available surface area
4
Freshly prepared samples of n-hexane or toluene soot were
deposited directly on the sample holder by burning the vapor
of about 30 mL of n-hexane (EM Science, purity 98.5%) or
toluene (EM Science, purity 99.5%) in a small beaker. The
sample was placed at a height of ∼5 cm from the top of the
beaker so that soot is collected from the top part of the flame.
Particle characterization studies of hexane and toluene soot were
performed in order to have some insight about the nature of
these particles. These characterization studies include particle
size distribution analysis using transmission electron microscope
images, specific BET surface areas, SBET, and bulk density, Fb,
measurements. Figure 1 shows representative TEM images of
n-hexane soot and toluene soot. Table 2 lists the physical
parameters of hexane and toluene soot. It should be noted that
the size of the particles used here is close to the 30-50 nm
diameter carbon particles found in aircraft exhaust.22
difficult. As noted by Longfellow et al., in order to apply
laboratory measurements of the heterogeneous uptake measure-
ments to the atmosphere, the surface area of the soot must be
considered.
Another difference in the literature values is that the initial
uptake coefficient is reported in some cases, whereas a steady
state or an average value is reported in others. This means that
the reactivity of a surface covered with adsorbed molecules is
compared to the reactivity of the unreacted surface. Because of
site-blocking, adsorbate-adsorbate interactions, and electronic
effects, uptake coefficients are typically coverage dependent and
usually decrease as a function of coverage.1
8,19
In this study, the reaction of NO2 with freshly prepared hexane
soot has been studied using a Knudsen cell reactor and FT-IR
spectroscopy. A limited number of experiments were done on
toluene soot as well. The uptake coefficient for the NO2‚soot
reaction was measured using a Knudsen cell reactor. Using
models that take gas diffusion into account, an uptake coefficient
that takes into consideration the accessible surface area and the
increased number of gas surface collisions within the underlying
layers is determined. We report here initial and average values
of the uptake coefficient in order to compare to literature values.
In addition, we have quantified gas-phase reactants and products
in order to determine the total amount of NO2 reacted with the
soot surface as well as the branching of NO2 uptake on the
hexane soot surface to the production of HONO. FT-IR
experiments were done in order to gain some additional insight
into the heterogeneous reaction of NO2 on soot and the possible
For the Knudsen cell experiments, soot samples were evacu-
-
8
ated overnight to reach a pressure of ∼5 × 10 Torr inside
the reactor. The sample holder was then sealed and NO2
(
Matheson, 99.95% purity, used as received) was introduced to
the reaction chamber through a leak valve to passivate the walls
of the reactor. Experiments were run at NO2 pressures near 8
µTorr (equivalent to 11 ppb or 2.5 × 10 molecules cm ).
The gas pressure inside the reaction chamber was monitored
by an absolute pressure transducer (MKS 690 A.1TRC, range
.1-10 Torr). All experiments were done at 295 K.
Prior to each experiment, calibration of pressure vs mass
spectral intensity data for pure gaseous NO2 (m/e ) 46) was
made. By using pure NO2 for this calibration, the contribution
1
1
-3
-
6
0
4,6
role of adsorbed water that has been reported for this reaction.
15
of NO2 (m/e ) 47) to the m/e ) 47 ion signal monitored during
the course of the experiment can be calculated. The calibration
of pure NO2 also provides information about the fragmentation
of NO2 to NO (m/e ) 30). Calibration of HONO is more
difficult, since there is no readily available source of pure
gaseous HONO. The mass spectrum of HONO shows a parent
ion peak at m/e ) 47 and two major fragments at m/e ) 17
Experimental Section
A Knudsen cell reactor coupled to a quadruple mass
spectrometer was used to study the kinetics of the uptake of
NO2 on hexane and toluene soot and to quantify gas-phase
reactants and products. The reactor consists of a stainless steel
cross that serves as the reaction chamber. The region between
the reaction chamber and the mass spectrometer is separated
by a gate valve and the escape aperture or escape hole. The
size of the aperture can be changed in a vacuum using a linear-
rotary motion feedthrough. The quadruple mass spectrometer
+
+
(OH ) and m/e ) 30 (NO ); there is no apparent fragment at
+
23
m/e ) 46 (NO2 ). However, the parent ion peak (m/e ) 47)
15
has two contributions, the mass spectral intensity of NO2 and
1
5
HONO. Since the contribution of NO2 to the m/e ) 47 signal
1
4
is known from the NO2-signal, the mass spectral intensity of
HONO can be determined. In addition, calibration of pure
gaseous NO (Matheson, 99%) pressure versus mass spectral data
of m/e ) 30 ion signal was also done. The production of NO
can be calculated by correcting for contributions to the ion
(
UTI, 100C) is pumped by a 150 L/s ion pump, and the region
between the reactor and the mass spectrometer is pumped by a
0 L/s turbo pump (both from Varian). The stainless steel sample
holder sits on top of a tee support in the reaction chamber and
7