INTRODUCTION
T lymphocytes (T cells) form an essential part of the adaptive immune system and are therefore of major
interest for the research community. Two recent examples are the work of the international ONE Study
(
www.onestudy.org) and BIO-DrIM (www.biodrim.eu) consortia, international groups of experts in the field
of immune monitoring, funded by the European Commission. Within these approaches, marker and dye
selections for flow cytometry have been designed and optimized by expert flow labs to monitor the human
immune response [1].
The close cooperation between these groups and Beckman Coulter resulted in the development of the
DuraClone IM brand of dry reagents for the analysis of human immune cells. Using the 10 color DuraClone
IM T Cell Subsets kit plus 3 additional liquid markers, a 13 color tube has been designed for the phenotypic
analysis of major T cell sub-populations in the human peripheral blood.
PROTOCOL
Standard Procedure:
After informed consent, 100 µL of human peripheral blood from a healthy donor was added to a DuraClone
IM T Cell Subset dry reagent tube (Beckman Coulter), followed by 5 µL each of Brilliant Violet 605TM
anti-human CD95, Brilliant Violet 650TM anti-human CD25, and Brilliant Violet 785TM anti-human CD127
antibodies (BioLegend). Cells were mixed for 8 seconds, incubated for 15 minutes at room temperature
(
RT) in the dark, and red blood cells were lysed by adding 2 mL of VersaLyse Lysing solution plus 50 µL of
IOTest 3 Fixative Solution (both from Beckman Coulter). Following incubation (20 min at RT), the suspension
was spun down (200 x g, 5 min), the supernatant discarded, and the pellet resuspended in 3 mL 1 x PBS.
After an additional centrifugation step (see above), the cell pellet was resuspended in 500 µL 1 x PBS for
subsequent analysis on a 13 color / 3 laser CytoFLEX flow cytometry system (Beckman Coulter).
CytoFLEX is the first commercial flow cytometer to utilize fiber array photo diodes (FAPD)s for fluorescence
channel detection. The FAPD provides low-noise detection with high quantum efficiency and minimum light
loss ensuring high signal to noise ratio and optical resolution especially with small particle measurements and
dim fluorescence detection.