3. Results and Discussion
gave a 56% yield with reduced selectivity. However a similar
run (Run 21) with halved concentrations gave 58% yield with
high selectivity.
The complete set of results for reactions (runs 1-22) utilising
the Kenics static mixer system as described above (arranged
chronologically), and comparable batch reactions (runs A-F)
mixed by adding the NaOCl solution to the TNT solution in a
stirred reactor, are presented in Table 3.
Runs 1-11 were essentially exploratory in nature to ascertain
the proper functioning of the system and to make any adjust-
ments that were deemed necessary.
From previous work10,13 it is known that the byproduct
R-chloro-2,2′,4,4′,6,6′-hexanitrobibenzyl (R-chloro·HNBB) in
crude HNS is converted by recrystallisation from NMP into
2-chloro-2′,4,4′,6,6′-pentanitrostilbene, which remains in the
recrystallised HNS. It is therefore important to reduce the
production of this byproduct as much as possible, even at the
expense of increasing the HNBB content since the latter is lost
on recrystallisation. In the present flow runs, the isolated HNS
contained 4-8 mol % of this byproduct. Other results from
batch experiments10,13 indicate that a higher pH level during
the after-reaction period favors the formation of R-chloro·HNBB.
Attempts to synthesise hexanitrobibenzyl (HNBB), by
reducing the equivalents of NaOCl to ∼0.5, failed due to
unreacted TNT separating from the reaction mixture (due to
the increased water content after mixing). This blocked the static
mixer (Runs 4 and 8).
Runs A and B were normal batch reactions (essentially
commercial conditions) with different equivalents of NaOCl
(1.36 and 1.00 equiv, respectively). The yields of crude HNS
were 36 and 44%, respectively, the HNS contents were 93 and
87%, respectively, and the HNBB contents were 1.5 and 6.4%,
respectively. As observed previously (see Introduction 1.1) the
proportion of HNBB decreases as the number of NaOCl
equivalents is increased.
Runs 3, 11, and 14 were quenched reactions performed to
measure the extent of reaction before the reaction mixture left
the exit tube. In Run 3 the exit tube was ∼3 m long; in Runs
11 and 14 it was ∼9 m long. From the amount of TNT
remaining in the quenched product, it was clear that the 9 m
exit tube was necessary for an almost complete reaction. With
only a 3 m exit tube, 30% TNT remained unreacted. If a
significant proportion of the TNT was still present at this point,
its continued reaction with NaOCl would generate heat in the
ageing vessel, giving less control over the after-reaction
conditions. This was observed as a more rapid temperature rise
(22 °C vs 13.5 °C after 30 min, respectively) in the collection
vessel in later runs. Run 11 indicated that moderately pure
TNBCl could be synthesised by this method, the only impurity
being TNT.
Run F was a standard (10-g scale) Shipp-Kaplan HNS
synthesis,3 and Run 22 was a static mixer run utilising the same
reactant concentrations and proportions as those in the
Shipp-Kaplan synthesis. Whilst the selectivity was high in both
cases, the yield of crude HNS was lower in the flow system
(28 vs 35%).
Runs 12 and 13 used the same number of equivalents of
NaOCl (∼1.3), but Run 13 used halved concentrations of both
reactants. The yield (37 vs 43%, respectively) and selectivity
(86 vs 94 mol % HNS, respectively) were better in the diluted
system.
Further observations relate to the observed temperatures, at
or after mixing. With water being pumped through both inlet
systems the observed temperature at the mixing point was 3.5 °C
(pump settings 40% and 27.5%). Thus, even though both liquids
had passed through 2 m of PTFE tubing immersed in ice-water,
the temperature of the liquid had only been reduced from
ambient to 3.5 °C. It is suggested that SS tubing, with its better
thermal conductivity, should be substituted for PTFE in the inlet
tubing coils in order to achieve better cooling of the inlet
solutions. That this temperature is the same as that observed
during an actual HNS run suggests that the present positioning
of the temperature sensor does not record the temperature
generated by the exothermic reaction on mixing. A temperature
sensor positioned at the bottom of the static mixing tube should
be explored.
The temperature in the collecting vessel during an actual
HNS run was never lower than 5.5 °C. This suggests that the
exothermic reaction may not be complete when the reaction
solution leaves the exit tube. This is supported by the observa-
tion in Runs 11 and 14 that some TNT remains unreacted at
this stage. Replacement of the PTFE exit tubing by SS tubing
should enhance the efficiency of cooling the reaction mixture.
This, of course, would reduce the rate of reaction further,
necessitating an extension (or broadening) of the exit tube in
order to achieve the same or increased extent of reaction at
this stage. Both changes would certainly be required if the pump
rate were to be increased in order to increase the scale of the
reaction.
Runs C and D were batch reactions run with pH control
(H2SO4, NaOH) during the after-reaction period, Run C at pH
10.2 and Run D at pH 10.4. Both used 1.0 equiv of NaOCl.
The yields of crude HNS were 57 and 62%, respectively. Flow
mixing, Run 17 (1.24 equiv of NaOCl) and Run 18 (1.02 equiv
of NaOCl), followed by control (H2SO4, NaOH) at pH 10.0
and 10.2, respectively, gave yields of ∼50% in both cases, but
selectivity was much higher with the greater proportion of
NaOCl (Run 17). A repeat of Run 18 with halved concentrations
(Run 19), gave essentially the same yield, but slightly better
selectivity.
In the present work, PTFE cooling coils were used both
before and after the mixer because they were more flexible to
work with. It was appreciated at the outset that SS coils would
give more efficient cooling. However, without experimental data
using the PTFE coils, it was not possible to say whether they
would have been of sufficient length to achieve the necessary
cooling.
Run E was a batch reaction run with pH control (pH 9.70)
using an ethanolamine buffer. The yield of crude HNS was 60%
with high selectivity. A corresponding static mixer run (Run
20), but with lower NaOCl (1.16 equiv; Run E used 1.36 equiv),
(13) Bellamy A. J. Unpublished work.
638
•
Vol. 14, No. 3, 2010 / Organic Process Research & Development