Communications
use of injection valves on the inlet lines offers the possibility
of using a wide range of substrates to efficiently produce a
library of products. Furthermore, with the microreactor
system, the reaction conditions themselves are improved.
The significantly greater mass-transfer area resulting from
segmented gas–liquid flow enables very rapid reaction times
from the accelerated mass transfer.[18]
The carbonylation case study demonstrates the consid-
erable potential of continuous-flow, microreactor-based
experiments at conditions not easily achieved in conventional
benchtop experiments. In particular, the technique provides a
useful tool for quickly and safely scanning reaction conditions
and reagents. By using this technique, we were able to test
multiple aryl halides over a wide range of temperatures and
pressures much more rapidly than could be accomplished
with batch experiments. For instance, in the study of the
effects of pressure and temperature on the aminocarbonyla-
tion of 4-bromobenzonitrile (5) up to 36 samples were
collected and analyzed in a single day. Table 1 summarizes
the conditions at which the maximum amide and a-ketoamide
yields for each reaction were observed. These results reveal
the general trend for an increase in the yield of amide with an
increase in temperature and an increase in the selectivity of a-
ketoamide at lower temperatures and higher pressures.
Figure 3. Product ratio of a-ketoamide (7) to amide (6) for the
aminocarbonylation of 4-bromobenzonitrile (5). Each data point repre-
sents an individual experiment.
installing injection valves onto the liquid reagent inlet lines.
Thus, if one desired to run several combinations of substrates
under the same conditions, this could be accomplished by
simply varying the solutions loaded into the inlet-sample
loops. This ability to rapidly change reactants and conditions
is a powerful strategy for fast synthesis of a diverse array of
compounds as well as for catalyst screening.
We demonstrated this ability by switching between 4-
bromobenzonitrile and 4-bromoanisole without stopping the
flow into the reactor. The system was brought to pressure with
toluene loaded in each syringe and the sample loops were
loaded with reagent stock solutions: Pd(OAc)2, Xantphos,
and toluene in one and Ar-Br, DBU (1,8-diazabicyclo-
[5.4.0]undec-7-ene), dodecane (internal standard), and mor-
pholine in the other. This system was successfully used to test
aminocarbonylation of 4-bromoanisole (Table 1, entry 5).
The use of a pressurized microreactor system greatly
expands the range of reaction conditions available to the
bench chemist. In this study, pressures from 4.5 to 14.8 bar
and temperatures from 98 to 1608C were examined with
greater flexibility, in terms of loading and sampling, than
would be possible with traditional high-pressure chemical
equipment such as a Parr bomb or autoclave. In addition, the
Experimental Section
Aminocarbonylation with carbon monoxide gas delivery directly
from gas tank: The system pressure was controlled by three needle
valves (Upchurch P-445). The cylinder outlet was split and each
branch was connected to a needle valve. One branch was connected to
the microreactor gas inlet and the second branch was connected to the
pressure bomb (pressurized collection vessel) makeup inlet
(Figure 1). The third needle valve was connected to the bomb
outlet tubing such that there was a controlled constant leak from the
headspace of the pressure bomb.
After the bomb was pressurized, all needle valves were closed.
The cylinder delivery pressure was then raised an additional 5% to
compensate for pressure drops in the delivery system. The liquid
reagent flow rate was next set to the desired reaction conditions, and
the reactor gas inlet was partially opened to allow gas flow and begin
slug flow equilibration. Additionally, the leak valve was slightly
opened to allow the microsystem to come to steady state. During
operation, these two valves were used to control the flow rate as
observed in the reactor by measuring the speed of the slugs. The
overall flow rate was varied between 30 and 150 mLminÀ1, which
corresponds to reaction times of 13–2.5 min, respectively. The bomb
inlet valve remained fully closed unless the pressure in the bomb
dropped below the desired operating pressure, in which case the valve
was opened to allow repressurization.
Table 1: Maximum yieldsfor variouscarbonylation reactions.
1
P [bar] T [8C] Av. time Conv. Yield P1 Yield P2
[min.]
[%]
[%]
Sample analysis: The samples were analyzed by gas chromatog-
raphy with an Agilent 6890 Series gas chromatograph and an FID
detector. The samples were injected by an Agilent 7683 automatic
liquid sampler into a 10-meter Agilent HP-1 capillary column
(internal diameter 200 mm; film thickness 0.11 mm) with a flow rate
of nitrogen 1 mLminÀ1 . The oven temperature was raised from 70 to
2408C over 6.5 minutes. Sample peak areas were normalized to the
response of the internal standard to determine the sample concen-
trations. Compounds were isolated by chromatography on the residue
7.9
7.9
146
116
3.3
4.2
100
100
68
35
28
65
2.7
14.8
160
109
7.1
6.6
100
99
83
32
0
57
2.7
150
12.7
48
35
0
1736
ꢀ 2007 Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim
Angew. Chem. Int. Ed. 2007, 46, 1734 –1737